Level Probe LGC2
Installation
2017
-0
2
27
• To make installation easier, Fuchs marks the extension cable if a customer
specific length has been ordered.
Ordering information: product structure, feature "Service", option "RI" or "SI".
• Cable marking tolerance (distance to lower end of level probe):
Cable length < 5 m (16 foot):
±
17.5 mm (0.69 inch)
Cable length > 5 m (16 foot):
±
0.2 %
• Material: PET, stick-on label: acrylic
• Immunity to temperature change: -30 to +100
°
C (-22 to +212
°
F)
5.10
Cable Shortening Kit
Figure 5.9
The cable shortening kit is used to shorten a cable easily and professionally. Ordering
information: product structure, feature "Accessories", option "WP"
5.11
Post-installation check
• Is the device undamaged (visual inspection)?
• Does the device conform to the measuring point specifications?
For example:
–
Process temperature
–
Process pressure
–
Ambient temperature
–
Measuring range
• Are the measuring point identification and labeling correct (visual inspection)?
• Are the securing screws tightened securely?
Note!
The marking is used exclusively for installation purposes. The mark must be thoroughly
removed without trace in the case of devices with drinking water approval. The extension cable
must not be damaged in the process.
Note!
Not for use of the device in hazardous areas.
Note!
The cable shortening kit is not designed for the device with FM/CSA approval.