Mounting the Temperature Head Transmitter
1. Guide the mounting screws (1) with the mounting springs (2) through the bore of the
temperature head transmitter (3).
2. Secure the mounting screws with the circlips (4). Circlips, mounting screws and springs
are included in the scope of delivery for the temperature head transmitter.
3. Screw the temperature head transmitter into the field housing tightly. (Width of screwdriver
blade max. 6 mm 0.24 inch)).
Figure 5.6
Warning!
Property damage from wrong mounting
If you tighten the mounting screws too tightly, this can damage the temperature head
transmitter.
Do not overtighten the mounting screw too.
1
Terminal box
2
Terminal block/terminals
3
Temperature head transmitter HUT
Measuring unit mm (inch)
> 7 (0.28)
1
2
3
Warning!
Property damage from wrong mounting
If the distance between the terminal block and the temperature head transmitter is too small,
this can result in damage of the components.
Maintain a distance of > 7 mm (> 0.28 inch) between the terminal block and the HUT
temperature head transmitter.