3.9. Linear Frequency
Range
AC Sensitivity (TRMS, sine wave )
5 Hz- 10 kHz
10 kHz-100 kHz
220 mV
> 10 mV
> 20 mV
2.2 V
> 100 mV
> 200 mV
22 V
> 1 V
> 2 V
220 V
> 10 V
> 20 V
600 V
>100 V*
Not specified
220 µA
> 10 µA*
2200 mA
> 100 µA*
22 mA
> 1 mA*
220 mA
> 10 mA*
10 A
> 1 A*
* Frequency Response: 50 Hz
– 10 kHz
3.10. Duty Ratio
Frequency range
Duty ratio range
Resolution
Accuracy
20 Hz ~ 10 kHz
5% ~ 95%
0,01%
± (10%)
3.11. Continuity Test
Range
Audible Threshold
Response Time
Test current
220
Less than 30
approx. 100 ms
< 0,7 mA
3.12. Diode Test
Range
Description
The approximate forward voltage drop of the diode will be displayed.
If the voltage drop is more than 2V, the display shows “OL”
3.13.Temperature Measurement
Temperature Range
Resolution
Accuracy
°C
-20°C - 0°C
0,1°C
± (6.0% + 3°C)
0°C - 400°C
± (1.5% + 3°C)
400°C - 1000°C
± (1.8% + 3°C)
°F
-4°F - 32°F
0,1°F
± (6.0% + 6°F)
32°F - 752°F
± (1.5% + 6°F)
752°F - 1832°F
± (1.8% + 6°F)
Using of e K type thermocouple
Note:
1. The accuracy does not include error of the thermocouple probe.
2. Accuracies apply after 1.5 hours if ambient temperature changes
-27-