-68-
8 Technical Specifications
8.1 Thermal Imager
Field of view (FOV) / Minimum focus distance
21° x 21°/ 0.5m
Spatial resolution (IFOV)
4.53mrad
IR resolution
80 × 80 pixels
Thermal sensitivity/NETD
< 0.1°C @ +30°C (+86°F) / 100 mK
Image frequency
50Hz
Focus mode
Focus free
Focal length
7.5mm
Focal Plane Array (FPA)
/
Spectral range
Uncooled microbolometer / 8
–14 µm
Object temperature range
–20°C to +260°C (–4°F to +500°F)
Accuracy
±3°C (±5.4°F) or ±3% of reading (Environment temperature
10
℃
-35°C
,
object temperature >0°C. )
Accuracy calculated as [%r (num. digits*resolution)] at 18°C ÷ 28°C <75%HR
DC Voltage
Range
Resolution
Accuracy
Input impedance
Protection against overcharge
600.0mV
0.1mV
±(0.09%r 5digits)
>10MΩ
1000VDC/ACrms
6.000V
0.001V
60.00V
0.01V
600.0V
0.1V
±(0.2%r 5digits)
1000V
1V
AC TRMS Voltage
Range
Resolution
Accuracy
(*)
Protection against overcharge
(50Hz
÷
60Hz)
(61Hz
÷
1kHz)
6.000V
0.001V
±(0.8%5digits)
±(2.4%5dgt)
1000VDC/ACrms
60.00V
0.01V
600.0V
0.1V
1000V
1V
(*) Accuracy specified from 10% to 100% of the measuring range,
sine wave. Input impedance: >9M
Ω
;
Accuracy PEAK function: ±10%rdg, PEAK response time: 1ms
AC+ DC TRMS Voltage
Range
Resolution
Accuracy (50Hz÷1kHz)
Input impedance
Protection against overcharge
6.000V
0.001V
±(2.4%r20dgt)
>10MΩ
1000VDC/ACrms
60.00V
0.01V
600.0V
0.1V
1000V
1V