Peak Atlas DCA Pro
User Guide
June 2019
–
Rev 1.7
Page 43
Specifications continued...
Parameter
Min
Nom
Max
Units
Notes
JFETs
Pinch-off VGS(OFF) range
-10
2.5
Drain current at Pinch-off
2.5
5.0
10.0
μ
A
Drain current at VGS(ON)
4.75
5.00
5.25
mA
Drain-source voltage at VGS(ON)
3.5
9.0
V
VDS for IDSS measurement
3.0
3.25
V
IDSS measurement range
1.0
12.0
mA
VDS for IDSS measurement
0.0
3.0
3.5
V
IDSS accuracy
±1.7% ±13.1
μ
A
RDS(ON) drain test current
4.75
5.00
5.25
mA
11
RDS(ON) gate-source test voltage
7.80
8.00
8.20
V
12
RDS(ON) accuracy
±2% ±2
Ω
RDS(ON) resolution
1.0
2.0
Ω
Transconductance range (gfs)
99
mA/V
Transconductance accuracy
±5% ±2mA/V
±10% ±5mA/V
Thyristors (SCRs) and Triacs
Gate test current
8.0
10.0
12.0
mA
6
Load test current
10.0
15.0
mA
Diodes and LEDs
Forward test current
4.75
5.00
5.25
mA
Forward voltage accuracy
±1.2% ±19mV
Acceptable VF for a diode at 5mA
0.15
V
VF for a LED decision at 5mA
1.50
4.00
V
Zener Diodes
Zener voltage range at 5mA
1.8
7.0
V
Zener voltage range below 5mA
6.0
11.0
V
Zener voltage accuracy
±1.2% ±19mV
Zener test current range
0.50
5.00
5.25
mA
Voltage Regulators
Input test voltage range (IQ<2.0mA)
1.1
10.0
V
Input test voltage range (IQ<4.0mA)
8.0
V
Acceptable quiescent current range (IQ)
0.0
6.0
mA
Quiescent current accuracy
±1.7% ±13.1
μ
A
Dropout voltage range
0.00
3.00
V
Output voltage accuracy
±1.2% ±19mV
CS344.550
Notes:
1. Collector current of 5mA. Gain accuracy for gains less than 2000.
2. Resistance across reverse biased base-emitter > 60k
Ω
.
3. Resistance across reverse biased base-emitter < 60k
Ω
.
4. Load current of 5mA.
5. Collector-emitter voltage of 5.0V.
6. Thyristor quadrant I, Triac quadrants I and III.
7. BJT with no shunt resistors.
8. Load current typically 10mA.
9. Transistor has leakage of more than 10
μ
A.
10. IGBT/MOSFET gate test voltage is reduced to 5V if gate leakage is detected.
11. Lower drain test current is used if IDSS<7mA.
12. For normally-off JFETs, a gate voltage that results in 10
μ
A of gate current will be used.