- PCME QAL 181 MANUAL: 73 -
Purge Flow
15 (adjustable)
50
Disabling Self Tests
To enable/disable the running of the zero/span checks:
Select ‘Setup’ display
Menu route:
More Settings
↵
Edit an existing device
↵
Select device to edit
Self Tests Active
Yes / No
Select Save to save the settings and return to the main menu.
Note this does not disable the
Purge Flow alarm.
To enable/disable the zero/span alarms:
Select ‘Setup’ display
Menu route:
More Settings
↵
Edit an existing device
↵
Select device to edit
Zero/Span alarm?
Yes / No
To adjust the Low Purge Flow alarm level:
Select ‘Setup’ display
Menu route:
More Settings
↵
Edit an existing device
↵
Select device to edit
Purge Alarm Level
The default level is 15 Litres/minute. To prevent the Low Purge Flow alarm, set the Purge Alarm
Level = 0.
Select Save to save the settings and return to the main menu.