- PCME QAL 181 MANUAL: 51 -
A Master Reset will delete all settings and logged data. This data should be
downloaded to a PC using the PCME PC software before carrying out a
Master Reset.
To perform a Master Reset
Select ‘Setup’ display
Menu route:
Other Functions
↵
Reset Functions
↵
Master Reset
The Communications Mode is set by PCME prior to installation and should have been checked
during the system installation.
To check the Communications Mode
Select ‘Setup’ display
Menu route:
Other Functions
↵
Advanced Functions
↵
Sensor Comms Mode
The settings should be as follows:
Setting
Mode
Baud Rate
Parity
Standard
ASCII
19200
Odd
Optional
RTU
19200
None
If problems are encountered communicating with a sensor, check that the sensor communication
setting (SW101) is the same as the multicontroller Sensor Comms Mode setting.
6.4
Autodetect Sensors
Sensor channels can be added to the multicontroller by autodetect or manually.
The manual mode should only be used to add sensors when manual control of the
order in which the devices are displayed is required or only specific sensors are to be
added.
In the ‘Autodetect’ mode the multicontroller scans all the modbus addresses to identify all the
sensors in the network.