Section 4 Continued…

 

 
7. RELAY FUNCTI ONS 

The unit may be user configured for several relay options. 
Ensure that the necessary hardware options are installed i.e. 1 or 2 relays. The relays are solid state 
devices. These relays are rated at a maximum of 100mA at 24vdc. Connections to the relays are via 
terminal connections close to the devices designated 1 & 2 =  relay 1 contacts 3 & 4 =  relay 2 contacts. 
These relays are volt-free contacts unless specified 
Three modes of relay control are selectable along with 3 patterns of operation as detailed. 

 
a. Standard Batcher (MENU> WEI GHMODE< code 4: [ 4000x] )

 

This allows a batching sequence utilising a single relay control.  
Press the MODE key the display will briefly prompt TRIP1” and then the existing value will be shown. To 
clear existing data first press the ZERO key to clear the value, the far left digit will show “E” indicating that 
the display is in keyboard mode. Press the PRI NT key to increment the least significant digit and press the 
TARE key to shift the digit to the left i.e. to input a value of 4000 press the PRI NT key 4 times and then 
press the TARE key 3 times.  
Repeat the above if Trip 2 is installed. 
Ensure that the indicator shows zero by selecting the ZERO or TARE key.  
Press the PRI NT key to initiate the sequence.  
The relay will close and remain closed until the target is reached. I f a tare is selected this will be cancelled, 
further batching is inhibited until the display indicates zero. Use of the TARE key to zero the display is 
intended for silo batching operations from a loaded silo. 
 

b. One trip batcher mode (MENU> WEI GHMODE< code 5 – AUTOMATI C: [ 5000x] ) . 

This mode controls a single relay continuously. 
 

c. Tw o trip batcher mode ( MENU> WEI GHMODE< code 6 – AUTOMATI C: [ 6000x] ) . 

This mode controls two relays continuously. 
 

RELAY PATTERN SELECTI ON 

Refer to the menu code table and menu setting instructions. 
3 relay patterns are available and each may include fast or normal integrate speeds.  

MENU> FUNCTI ON< code 0/ 1

 will operate normally 

open

 contacts, the relays will 

close

 if equal too or 

greater than trip values. 

MENU> FUNCTI ON< code 2/ 3

 will operate normally 

closed

 then 

open 

 if equal too or above the trip 

values. 

MENU> FUNCTI ON< code 4/ 5

 is intended for a low/ high system. Relay 1 is normally 

closed

  below its trip 

value then 

open

 above its trip value and relay 2 is normally 

open

 below trip value and 

closed

 at the preset 

high level trip value.

 

 
8. A.E.O. ( Analogue Expansion Option) : Menu select [00400] 

This menu setting relates to the Analogue Expansion Card that, when fitted, will transmit 4-20mA, 0-5vdc or 
0-10vdc @ 16bits resolution. This output is an analogue value of the displayed weight. Refer to the data 
sheet supplied with the unit for more details. 
 

Correct Procedure to charge unit: 

1. With the adaptor 

AND

 Junior switched 

OFF

 plug the adaptor jack into the unit. 

2. Switch on the Adaptor and allow approx 12 hours to fully charge (from a fully discharged state). You can 
use the unit whilst in charging mode. 
3. Once the charge cycle is complete press and hold the ZERO key to switch off the Junior and 

sw itch OFF 

the adaptor via the mains outlet.

 

4. Disconnect the Adaptor jack from the Junior and use normally until the Battery Low indicator is indicated 
then complete cycle. 

 

 

Содержание Apex Junior

Страница 1: ...1 Apex Junior Cased Weighing Indicator Controller Document Version JG8 101215 Software Version H Board revision 1 4 onwards APPROVED DISTRIBUTORS OF Electronic Weighing Services USER MANUAL...

Страница 2: ...ve 6 26v d c E Ground ii DC Applications Battery Versions B Battery Positive B Battery Negative V Adaptor DC 12VDC V Adaptor Earth Suitable Charging Adaptor approx 12vdc 300mA minimum Correct Procedur...

Страница 3: ...t manual G8 WI NDHT 120603 02 1 2 3 4 5 6 DP K G 0A O D 02 Start 6 digits weight plus placed decimal point data MSD to LSD Units i e kg g or te 0A line feed 0D Carriage return D2 SETTI NG UP SERI AL C...

Страница 4: ...The display will briefly prompt LOAD and then indicate E00000 awaiting the input of your test load value at this point check that the platform is empty then enter your required test load value via th...

Страница 5: ...E key note the unit must have the necessary hardware options installed if printing or relay functions are selected Weigh Mode Print Mode Baud Select Line Feeds Function 0 Standard 0 Nett Kg 0 1200 0 1...

Страница 6: ...O and then MODE key together Release the ZERO key and then the MODE key 4 PEAK HOLD To select peak hold mode press the MODE key Peak weight will then be displayed until the MODE key is pressed again t...

Страница 7: ...ching operations from a loaded silo b One trip batcher mode MENU WEI GHMODE code 5 AUTOMATI C 5000x This mode controls a single relay continuously c Two trip batcher mode MENU WEI GHMODE code 6 AUTOMA...

Страница 8: ...ad cells Conversion 24bit Sigma Delta A D I nput Range 0 150mV Linearity 0 0015 F S Temperature Drift 2 p p m per degree C Update Speed 7ips or 30ips user selectable I nternal Resolution 300 000 count...

Страница 9: ...E DESCRIPTION EN55022 Radiated Emissions EN61000 4 4 Fast Burst Transient EN61000 4 3 Radiated Immunity EN61000 4 6 Conducted Immunity EN61000 4 2 Electrostatic Discharge 2014 35 EU Low Voltage Safety...

Отзывы: