4
Section 2
CALI BRATI ON.
Press and hold the ZERO key and then press the POWER key, when the display indicate
“888888” release the ZERO key and the display will prompt ‘DP 0000’ and the last
decimal point setting. To alter the decimal point setting press the PRI NT key until the
required position is indicated, once selected press MODE to accept.
The display will now prompt "DI V" and the last minor division increment will be
indicated. To alter this value (possible selections are: 1, 2, 5, 10, 20 or 50) repeatedly
press the PRI NT key until the required value is shown, once selected press MODE to
accept
The display will then briefly prompt “TOP” followed by the last top capacity stored. The
far left digit will indicate “E” showing that the display is in keyboard mode. To clear
existing data press the ZERO key. Press the PRI NT key to increment the least significant
digit and press the TARE key to shift the digit to the left i.e. to input a value of 4000
press the PRI NT key 4 times and then press the TARE key 3 times. When you have input
the required data press MODE to store the value.
The display will briefly prompt “LOAD” and then indicate ‘E00000’ awaiting the input of
your test load value, at this point check that the platform is empty then enter your
required test load value via the keyboard and press MODE to accept. The display will
now auto null any dead load and raw un-calibrated weight will be indicated. Check that
the display is zero and press the ZERO key if necessary then position the test load onto
the platform, when the display is stable press the MODE key. The display will then
indicate the existing function code for the unit, the MODE key will accept this code or
refer to Menu settings if you wish to amend it. The unit is now calibrated and ready for
use.
Post Trim Calibration
This mode is useful for Silo and large capacity applications.
Switch on the unit whilst holding the MODE key. The unit will work as a standard
weighing machine with no functions. Press the TARE key to increase the displayed
weight value or the PRI NT key to decrease the displayed weight value. Once you have
the correct weight displayed press the MODE key to return to normal weighing mode.
LOADCELL ERROR MESSAGE
I f the display prompts “L-CELL” then the input voltage is outside the working range of
the unit. Check all load cell connections are present and correct.