Section 2  

 

CALI BRATI ON. 

 
Press and hold the ZERO key and then press the POWER key, when the display indicate 
“888888” release the ZERO key and the display will prompt ‘DP 0000’ and the last 
decimal point setting. To alter the decimal point setting press the PRI NT key until the 
required position is indicated, once selected press MODE to accept. 
 
The display will now prompt "DI V" and the last minor division increment will be 
indicated. To alter this value (possible selections are: 1, 2, 5, 10, 20 or 50) repeatedly 
press the PRI NT key until the required value is shown, once selected press MODE to 
accept 
 
The display will then briefly prompt “TOP” followed by the last top capacity stored. The 
far left digit will indicate “E” showing that the display is in keyboard mode. To clear 
existing data press the ZERO key. Press the PRI NT key to increment the least significant 
digit and press the TARE key to shift the digit to the left i.e. to input a value of 4000 
press the PRI NT key 4 times and then press the TARE key 3 times. When you have input 
the required data press MODE to store the value. 
 
The display will briefly prompt “LOAD” and then indicate ‘E00000’ awaiting the input of 
your test load value, at this point check that the platform is empty then enter your 
required test load value via the keyboard and press MODE to accept. The display will 
now auto null any dead load and raw un-calibrated weight will be indicated. Check that 
the display is zero and press the ZERO key if necessary then position the test load onto 
the platform, when the display is stable press the MODE key. The display will then 
indicate the existing function code for the unit, the MODE key will accept this code or 
refer to Menu settings if you wish to amend it. The unit is now calibrated and ready for 
use. 
 

Post Trim Calibration 

This mode is useful for Silo and large capacity applications. 
Switch on the unit whilst holding the MODE key. The unit will work as a standard 
weighing machine with no functions. Press the TARE key to increase the displayed 
weight value or the PRI NT key to decrease the displayed weight value. Once you have 
the correct weight displayed press the MODE key to return to normal weighing mode. 
 

LOADCELL ERROR MESSAGE 

I f the display prompts “L-CELL” then the input voltage is outside the working range of 
the unit. Check all load cell connections are present and correct.  

 
 
 
 
 
 
 

Содержание Apex Junior

Страница 1: ...1 Apex Junior Cased Weighing Indicator Controller Document Version JG8 101215 Software Version H Board revision 1 4 onwards APPROVED DISTRIBUTORS OF Electronic Weighing Services USER MANUAL...

Страница 2: ...ve 6 26v d c E Ground ii DC Applications Battery Versions B Battery Positive B Battery Negative V Adaptor DC 12VDC V Adaptor Earth Suitable Charging Adaptor approx 12vdc 300mA minimum Correct Procedur...

Страница 3: ...t manual G8 WI NDHT 120603 02 1 2 3 4 5 6 DP K G 0A O D 02 Start 6 digits weight plus placed decimal point data MSD to LSD Units i e kg g or te 0A line feed 0D Carriage return D2 SETTI NG UP SERI AL C...

Страница 4: ...The display will briefly prompt LOAD and then indicate E00000 awaiting the input of your test load value at this point check that the platform is empty then enter your required test load value via th...

Страница 5: ...E key note the unit must have the necessary hardware options installed if printing or relay functions are selected Weigh Mode Print Mode Baud Select Line Feeds Function 0 Standard 0 Nett Kg 0 1200 0 1...

Страница 6: ...O and then MODE key together Release the ZERO key and then the MODE key 4 PEAK HOLD To select peak hold mode press the MODE key Peak weight will then be displayed until the MODE key is pressed again t...

Страница 7: ...ching operations from a loaded silo b One trip batcher mode MENU WEI GHMODE code 5 AUTOMATI C 5000x This mode controls a single relay continuously c Two trip batcher mode MENU WEI GHMODE code 6 AUTOMA...

Страница 8: ...ad cells Conversion 24bit Sigma Delta A D I nput Range 0 150mV Linearity 0 0015 F S Temperature Drift 2 p p m per degree C Update Speed 7ips or 30ips user selectable I nternal Resolution 300 000 count...

Страница 9: ...E DESCRIPTION EN55022 Radiated Emissions EN61000 4 4 Fast Burst Transient EN61000 4 3 Radiated Immunity EN61000 4 6 Conducted Immunity EN61000 4 2 Electrostatic Discharge 2014 35 EU Low Voltage Safety...

Отзывы: