RH5 Humidity Measuring Device

 

with 

S

word 

S

ensor 

Version 

1

.0_en 

© 

PCE Instruments

201

6

 

Содержание RH5

Страница 1: ...User Manual RH5 Humidity Measuring Device with Sword Sensor Version 1 0_en PCE Instruments 2016...

Страница 2: ...orm the sensor For heavy stacks and rolls please use the optionally available sword sensor holder and the tool for removing the sword sensor holder Measurement To switch on the instrument press the ke...

Страница 3: ...ative humidity the excessive moisture would condense or form fog Dew point temperature The dew point indicates the temperature that the not completely saturated air has to reach in order to be complet...

Страница 4: ...page 4 Design of the device batteries slot flipside measuring value calibration curve ON OFF button temperature selection of calibration curves air humidity and temperature sensor Option interface...

Страница 5: ...r ON OFF Switch upper Switch lower Save Hold Autolog Watch saved data Enter suppliers data Menu Enter Switch upper Switch lower Exit Enter numbers Enter letters Next or right Left Yes No Shift OK Edit...

Страница 6: ...ase press the button Supplier s data can also be entered on the PC subsequently Switching on the display lighting Press the key briefly the display lighting switches off automatically after approx 20...

Страница 7: ...until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs or Auto Logs and confirm with again All saved logs will be transferred to...

Страница 8: ...e cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please change the batteries and try again Press the button at your devi...

Страница 9: ...temperature of the different components is to store all components together in a room with only small temperature fluctuations minimum over night better for 24 hours Components of calibration equipme...

Страница 10: ...carefully and proceed as follows Conditioning the sensor To achieve best results let the sensor adjust for two hours The temperature has to be between 20 C and 26 C If the shown measuring value diffe...

Страница 11: ...factory calibration as follows Reset to factory calibration 1 Press the button two times to reach the menu point Options 2 Select the menu item Reset using the button and confirm by pressing 3 Enter...

Страница 12: ...ervals for checking the instrument depend on your operational demands and the required level of accuracy In general the drift of the sensor according to the degree of use constant humidity or use with...

Страница 13: ...emove the batteries For eventual resulting damages we cannot provide any warranty In case of a faulty device please contact PCE Instruments Optional accessories PC interface for printing saved data on...

Страница 14: ...page 14 Notes...

Страница 15: ...h German French Italian Spanish Russian 4 pcs 1 5Volt AA Alkaline batteries for approx 1800 measurements after approx 4 minutes 30 mA with display lighting 128 x 64 matrix display with LED backlightin...

Страница 16: ...ing To demonstrate the importance of temperature adjustment the table below shows measuring errors due to a temperature difference of only 1 C 1 8 F between the measuring instrument and the substance...