5. Enter the superuser password using the buttons
resp.
and confirm
by pressing .
•
The superuser password after consignment is the serial number of the
device, shown on the display after switching
on the device or in menu item
Status
.
6. A query if a setting is desired appears. Confirm by
pressing
.
7. Wait until the bar has risen completely. The
device adjusts by itself and automatically jumps
back to the measuring window. The adjustment is
completed now.
8.
Check the result before you remove the device
from the calibration equipment. Depending on the
temperature the display should show a humidity
around 50% now.
If you made a mistake during the setting, you can reset to the factory
calibration as follows:
Reset to factory calibration
1. Press the
button two times to reach the menu point
Options.
2. Select the menu item
Reset
using the button and confirm by pressing
.
3. Enter the superuser password using the buttons
resp.
and
confirm by pressing .
•
The query
reset?
appears on the display.
4. Press the button for resetting the device to the factory calibration.
•
The software reloads the factory calibration data and reboots the
device. This will need about 15 to 20 seconds.
5. Pressing the button you can exit without any changes.
Conditioning of the sensor
The conditioning of the sensor (time until the device shows the actual
measuring value) depends on several parameters. The parameter
responsible for the highest measuring error is a temperature discrepancy
between the sensors resp. the whole measuring instrument and the material
to measure resp. the air.