5. Now put the third gasket ring into the bottom part.
6. Fit the metal ring on the third gasket.
7. Take the upper part with the RH5 and attach these carefully at the
bottom part of the
calibration device.
8. Pick up the RH5 together with the calibration device STRAIGHT and
DON´T TURN IT AROUND. Screw it up like shown in the picture.
Then put the RH5 with the calibration device down on a table carefully and
proceed as follows:
Conditioning the sensor
To achieve best results, let the sensor adjust for two hours.
The temperature has to be between 20°C and 26°C.
If the shown measuring value differs more than the factory tolerance(1.5%
r.h.), we recommend to carry out a recalibration as follows.
Offset adjustment
1. Leave the measuring device in the calibration equipment, and switch it
on.
2. Press the Rolling Menu button until you reach the main menu.
3. Select the menu item
Options
by pressing the button
and confirm by
pressing
.
4. Navigate to
Ajust
using the button and confirm by pressing
again.