9
When tick box is checked, the
instrument will alert the user
when the surface temperature
is less than 3°C (5°F) above
the dew point temperature (
value). According to ISO 8502-
4 and ASTM D3276, steel surfaces should general-
ly be 3°C (5°F) above the dew point when painting.
The instrument will alert the user with an audio tone
and the LCD screen will reverse (see above) when
conditions may be unsuitable for painting.
Alarm
To conserve battery life, displayed values
only update at the Log Interval. The last recorded
values will remain “frozen” on the display until the
next interval values have been stored.
Recorded datasets are stored into the currently
opened batch. If no batch is open, a new batch will
be created. Datasets are recorded regardless of the
state of the Smart Trend Indicators (see pg. 4).
To turn off Auto Log, press the button and hold
until the next tick sound. A final dataset will be stored
in memory and the current batch will remain open.
Recorded datasets can be viewed on the LCD,
viewed graphically for quick trend analysis on-site
(see pg. 7) or downloaded to a printer or PC.
Downloading to a PC requires optional PosiSoft
Software (pg. 12).
-It is recommended that fresh alkaline batteries be
installed prior to an extended Auto Log.
-Auto Log works best with optional Magnetic Surface
Temperature kit (see pg. 11)
-If memory becomes full, or the instrument loses
power and shuts off, all previously stored readings
will remain in memory.
NOTES:
9
NOTE: