![PCE Instruments PCE-LOC 20 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/pce-instruments/pce-loc-20/pce-loc-20_user-manual_1516143040.webp)
© PCE Instruments
38
HART Settings
Select YES to add a Series resistor (250Ω) into the mA circuit.
You can then use this instrument together with a HART®
communicator to set up and calibrate HART® devices.
This option is applicable for mA(24V) Read Power Input Type
Only.
Display Settings
Display
Intensity
Display Brightness Settings.
Range:
1
… 100
Display Off
Time
Standby Time in second after which display
will turn Off. To turn the display off press any
key.
Range:
0
… 9999 sec
Setting 0 will disable this feature. That means
display will never turn off automatically.
Date/Time Settings
To set the Time + Date of the device
There are 2 Date formats supported in this device DD/MM/YY &
MM/DD/YY. This is useful only in Data Logging, to decide in
which format Date to be stored.
There are 2 Time formats supported in this device 24 Hour & 12
Hour. This is to select in which format the time should be
displayed on Run page & time to be stored in Data Logging.
AM/PM selection is enabled only for 12 Hour Time Format.
Содержание PCE-LOC 20