©  PCE Instruments 

14 

 

2. 

Measure Only 

 

 

Measure Window 

Input Type 
Measure Display Mode 
Measure Reading 
HART Icon 

See chart in 3.2.4 Display Mode 1 M Source 

Bar Graph 

Horizontal  Bar  graph  according  to  Input  Percentage  Value 
(0.00% 

… 100.00%). 

Percentage Value  

The Percentage Value in (0.00% 

… 100.00%)  according to 

Input Value. 

Tare 

The Tare Value Set from 

MENU 

 

DISPLAY 

 

MEASURE-

Tare 

page

 

 

Actual Value 

The Raw Input Value without any scaling 
 
This will appear only if 

Main Display

 in  

MENU 

 

DISPLAY 

 

MEASURE  

is set to 

PERCENTAGE/SCALED.

 

Min 

Displays  the  minimum  value  found  after  a  measurement  was 
started or minimum was reset. 

Max 

Displays the maximum value found after a measurement was 
started or maximum was reset. 

Max-Min 

Displays  the  Maximum-Minimum  value  found  after  a 
measurement was started or Maximum-Minimum was reset. 

Cumulative Average 

Displays  the  Cumulative  Average  value  found  after  a 
measurement was started or Cumulative Average was reset. 

 
 

 

Содержание PCE-LOC 20

Страница 1: ...Loop Calibrator PCE LOC 20 User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...attery 8 3 Start Up and Basic Operations 9 3 1 Power ON or OFF 9 3 2 User Interface 9 4 Menu Layout 20 4 1 MENU page 20 4 2 MEASURE Page 20 4 3 SOURCE Page 22 4 4 DISPLAY Page 27 4 5 DATA LOGGING Page...

Страница 3: ...ible damage If any damage is visible do not use the device Do not use the instrument in explosive atmospheres The measurement range as stated in the specifications must not be exceeded under any circu...

Страница 4: ...ument does not operate according to its specifications please contact PCE Instruments as soon as possible Delivery contents 1 x Calibrator PCE LOC 20 2 x Laboratory cables with 2 mm plug 2 x Adapter 2...

Страница 5: ...n of modules The configuration of your PCE LOC 20 may vary significantly from the one in the picture 1 Terminal Connection for Measure and Source 2 Keypad Section 3 Color Display 4 USB Connection Slot...

Страница 6: ...rent Source mA 2 Wire Simulation Voltage Generation The PCE LOC 20 is capable of voltage generation with two voltage generation ranges The following picture displays the connection for voltage source...

Страница 7: ...mA Current Measurement In this mode the PCE LOC 20 is not providing any supply voltage For proper measurement the external device should be capable of providing the voltage supply If the external devi...

Страница 8: ...in different menus These are shown in the bottom left part of the display This key has different functionalities in different menus These are shown in the bottom left part of the display Scroll down...

Страница 9: ...lied with the device is USB A type male to USB B type male It is used for connecting charger and PC Stand for Table Top Use This stand offers the best support for table top use which gives good viewin...

Страница 10: ...n you set the power on the battery symbol at the top of the display shows the charging status To get more information on the battery go to the Battery Info Page in the Setting menu Charging time Charg...

Страница 11: ...ower button until the display comes on During the power on sequence the instrument shows a startup message and then shows the applicable data To power OFF the instrument press and hold 2 seconds the s...

Страница 12: ...elements are included in the previous picture but the most important ones are discussed in the following chapters The Status Bar The status bar at the top of the display is visible only in RUN Page I...

Страница 13: ...formation 3 Measure mode not working 4 Device unable to get source feedback reading 5 Data log memory corrupt 6 Source mode not working 9 More than one error from above list is occurring 3 USB Connect...

Страница 14: ...e Input Measure Display Mode The Measure Reading Display Mode Actual Displays the Raw Input Value without any scaling Percentage Displays the Percentage Value in 0 00 100 00 Scaled Displays the Scaled...

Страница 15: ...e Actual Displays the Raw Output Value without any scaling Percentage Displays the Percentage Value in 0 00 100 00 Scaled Displays the Scaled Value Source Reading The Reading as per the Source Display...

Страница 16: ...e page Actual Value The Raw Input Value without any scaling This will appear only if Main Display in MENU DISPLAY MEASURE is set to PERCENTAGE SCALED Min Displays the minimum value found after a measu...

Страница 17: ...t in 3 2 4 Display Mode 1 Measure Source Bar Graph Horizontal Bar graph according to Output Percentage Value 0 00 100 00 Percentage Value The Percentage Value in 0 00 100 00 according to Output Value...

Страница 18: ...he Source Reading value after the switch OPEN was detected Switch CLOSE Reading Displays the Source Reading value after the switch CLOSE was detected Display Operations Mainly four types of widgets ar...

Страница 19: ...ns Select one of the options with the ENTER key Example How to change of Input Type I P Type from mA to mV This Option is available in MENU MEASURE Page ii EditBox EditBox is used where a large range...

Страница 20: ...changing of sign is not allowed But there are few EditBox where these are allowed These types of EditBox are Scaled Low 0 High 100 Range of measure source Alarm Low High value and Editing of Source Va...

Страница 21: ...tonBox is used where very few values are possible and all the available values need to be visible In this device two types of RadioButtonBox are available One with 1 value can be selectable and the ot...

Страница 22: ...Mode like Output Type Range Source Type etc DISPLAY Contains Parameters related to different display mode for RUN page LOGGING Contains Parameters related to Data Logging This ALARM Contains Paramete...

Страница 23: ...easure Input Range Input Range Low 0 to Default Input High This parameter is enabled if Main Display in MENU DISPLAY MEASURE is set to Percentage or Scaled Scaled Input Range Low 0 Scaling Low Range f...

Страница 24: ...OURCE The description of the parameters that appear on this page is given below Parameter Name Description Options O P Type Output Type Source Output Type Available Options mA 0 000 24 000 mA DC mA 2W...

Страница 25: ...splay in MENU DISPLAY SOURCE is set to Scaled Scaled Output Range High 100 Scaling High Range for Source Output Range Scaled Output Range Low 0 to 99999 Decimal Point for this EditBox can be changeabl...

Страница 26: ...ked Step unit Step Size in unit Step Size in unit where unit is changed according to current Output Type and Source Display Mode Only Specify one Step unit or Step the other will automatically changed...

Страница 27: ...e will increase or decrease Source Value by Step Size specified in STEP Page In RUN Page Source Value can directly change by Pressing F1 key EDIT and modifying value like in EditBox and STEP Setting c...

Страница 28: ...to Increase from Low to High Level Range 1 9999 Hold 100 s Time to wait at High 100 level in second This parameter is use for Repeat Format UP DOWN or DOWN UP Range 0 9999 Fall Time s Time to decrease...

Страница 29: ...ll change to PAUSE and STOP respectively So by pressing F1 and F2 key running RAMP can be PAUSE or STOP at any time in RUN Page 4 4 DISPLAY Page This Page appears when you select RUN MENU DISPLAY Ther...

Страница 30: ...MEASURE Scaled Display the Scaled Value of the Input The Scale Value depends on Input Range Input Scaled Range Transfer Function These settings are available from MENU MEASURE Filter sec 1st Order IIR...

Страница 31: ...a measurement was started or maximum was reset Max Min Displays the maximum minimum value found after a measurement was started or maximum minimum was reset Avg of 2 Displays the average value of pre...

Страница 32: ...set Maximum Displays the maximum value found after a measurement was started or maximum was reset Max Min Displays the maximum minimum value found after a measurement was started or maximum minimum wa...

Страница 33: ...le Value depends on Output Range Output Scaled Range Transfer Function These settings are available from MENU SOURCE Additional Info 1 Choose which information to be shown in Bottom Right side of the...

Страница 34: ...ose when Input Voltage Trigger Voltage Switch Open when Input Voltage Trigger Voltage All the MEASURE Page parameters are disabled if Switch Test Display Mode is selected Trigger Voltage Trigger Volta...

Страница 35: ...Description Options Trigger Data Logging Trigger Mode Selection Available Options Key Press Log Data on pressing from RUN key Page Periodic Log Data periodically at every Sampling Rate for total time...

Страница 36: ...rmation of stored files This information contains Logging Start Time Date and No of Samples stored in the file File Delete Delete stored file START LOGGING Press F1 key while selecting this button to...

Страница 37: ...s and a message LOGGING DONE pops up Transferring the Results to a Personal Computer A 32 bit Windows software called mCAL exe can be downloaded here https www pce instruments com english download win...

Страница 38: ...PCE Instruments 36...

Страница 39: ...rdance with Input Range and Measure Display Mode And also it can t be less than Alarm Low Limit when Low Alarm is enabled 4 7 SETTING Page This Page appears when you select RUN MENU SETTING All the av...

Страница 40: ...ay will turn Off To turn the display off press any key Range 0 9999 sec Setting 0 will disable this feature That means display will never turn off automatically Date Time Settings To set the Time Date...

Страница 41: ...shows the basic battery Information Battery Level Percentage Battery Voltage in mV Current in mA Battery Status Time to Full in min Time to Empty in min About Calibrator This Page illustrates the Con...

Страница 42: ...range Resolution Accuracy 0 250 mV 0 01 mV 0 02 of rdg 2 Dgt Voltage DC V 0 30V 0 001V 0 02 of rdg 2 Dgt Current DC mA 0 24 mA 0 001 mA 0 02 of rdg 2 Dgt Simulation parameters Voltage mV 0 250 mV 0 0...

Страница 43: ...Special features Step and ramp function Automatic and manual mode x x2 For the measuring function Continuity test Adjustable threshold up to 100 Power supply 3 7V 2300 mAh Li ion battery Charging tim...

Страница 44: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Страница 45: ...975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www...