background image

Manual 

10 

5.3.3 

Connection to a PC or printer 

To connect the device to a PC or printer, press the Menu button 

, select “Comm“ and press Enter 

Now, you will get to the communication screen. 

After that, use the USB cable to connect the device to a PC and press Enter 

. If the connection was 

successful, the screen shows “Communicating…” 
 

 

 
The device can also be connected to a printer. In this case, all measuring results are printed 
automatically. 

5.3.4 

Tolerances 

To set up or adjust tolerances, press the Menu button 

, select “Tolerance“ and press Enter 

. Now, 

you get to the tolerance screen. 

Here you can set up tolerances. To do so, use the arrow keys (

 and 

) to adjust the selected digit. 

Press Enter 

 to go to the next digit. After confirming the last digit by pressing Enter 

, the tolerance is 

set up. 

5.3.5 

Delete recorded data 

To delete the recorded data, press the Menu button 

, select “Delete“ and press Enter 

. Now, you get 

to the deletion screen. 
Here you have two different options. You can either delete all sample measurements (“Delete ALL 
Samples”) or all records (“Delete ALL Records”). 
Select the desired option by using the arrow keys (

 and 

) and press Enter 

. A confirmation screen 

appears. Press Enter 

 to confirm or Back 

 to cancel. 

 

Note: If you delete all records, all sample measurements will be deleted as well.

 

5.3.6 

Averaging 

You can set the number of measurements which are performed during the measuring procedure. From 
these particular measurement results, the average value is generated. 

To get to the averaging settings, press the Menu button 

, select “Average“ and press Enter 

Here, you can set the number of measurements. Use the arrow keys (

 and 

) to adjust the selected 

digit. Press Enter 

 to go to the next digit. After confirming the last digit by pressing Enter 

, the 

averaging is set. 
 

Note: If you set “00” or “01”, there will be no averaging. 
 

 

Содержание PCE-CSM 3

Страница 1: ...ter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Busines...

Страница 2: ...PC or printer 10 5 3 4 Tolerances 10 5 3 5 Delete recorded data 10 5 3 6 Averaging 10 5 3 7 Display mode 11 5 3 8 Language 11 6 Settings 11 6 1 Auto save 11 6 2 Time setting 12 6 3 Print settings 12...

Страница 3: ...suring device Please avoid strong changes in the environmental conditions This includes flickering lights as well as changes of the air temperature or air humidity The device may only be used in the a...

Страница 4: ...Lamp life 3 years more than 1 million measurements Storage 100 reference values standards 20 000 samples Interface USB RS 232 Power supply Rechargeable lithium ion battery 3 7 V at 3 200 mAh Charging...

Страница 5: ...Manual 5 4 System description 4 1 Button description 4 2 Interfaces...

Страница 6: ...USB interface you can transfer data from the colorimeter to a PC The baud rate is 115 200 bps 4 3 Battery Installing the battery 1 Make sure that the colorimeter is turned off 2 Open the cover of the...

Страница 7: ...easuring mode is L a b C h 5 2 Take a measurement If you turn on the device you are directed to the Standard Measurement screen To take a measurement follow these steps 1 Press and hold the testing bu...

Страница 8: ...took the standard measurement 3 Now you can see the results and the deviation to the standard in the following screen Note While the measurement is in progress time duration 1 second all buttons are i...

Страница 9: ...ation To get to the calibration menu press the Menu button select Calibrate and press Enter Here you can select between white and black calibration Use the arrow keys and to select and press Enter to...

Страница 10: ...ta To delete the recorded data press the Menu button select Delete and press Enter Now you get to the deletion screen Here you have two different options You can either delete all sample measurements...

Страница 11: ...nguage and press Enter Now you get to the language selection You can choose between English and Chinese Select an option by using the arrow keys and and press Enter to confirm 6 Settings To get to the...

Страница 12: ...he arrow keys and for selection and press Enter to confirm 6 3 Print settings Here you can choose if the measurement results are printed automatically or not when a printer is connected to the colorim...

Страница 13: ...alibration After that perform a black calibration 2 Fill in the powder Fill the powder into the powder compartment of the measuring plate After that put the powder test box back together see 7 1 1 Mak...

Страница 14: ...u can determine the colorimetric parameters of liquids pastes and powders 7 2 1 Build up The PCE CSM UTC is built up of a base plate 1 an alignment plate for the colorimeter 2 a measuring chamber 3 wi...

Страница 15: ...e measuring hole of the measuring chamber 3 Make sure that the colorimeter is in the right place and tightened properly before moving on 2 Fill the tank 4 with the desired medium and put it into the m...

Страница 16: ...mber Please make sure that these indications are aligned properly see picture below 4 Perform a measurement with the colorimeter 5 Clean the measuring device before each use Do not use abrasive or cor...

Страница 17: ...contact PCE Instruments 8 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 8 2...

Отзывы: