![PCE Health and Fitness PCE-TDS 100 H Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/pce-health-and-fitness/pce-tds-100-h/pce-tds-100-h_user-manual_1515713035.webp)
© PCE Instruments
31
M51
Set start time, interval and go on (run) time of the logger
M52
Select memory
1. To RS232
2. To Buffer
3. Buffer => RS232
The RS232 signal is transmitted via the USB interface.
M53
View saved data in internal memory; navigate through data using Reset, Back,
Up and Dow n keys; indication is updated automatically if logger has been
activated
M60
View date and time (calendar for 99 years); press ENTER to make changes;
press Reset to go to the next digit
M61
View software version and serial number (ESN)
M62
Setup USB interface:
Baud rate from 75 to 115200 bps
Parity None, Odd or Even
M67
Select frequency range for output betw een 0 and 9999, default setting is 1 to
1001 Hz
M68
Set volume flow for low est frequency
M69
Set volume flow for highest frequency
M70
Set display backlight; select how many seconds backlight w ill glow without
keystroke
M72
Reset w orking timer; press ENTER and confirm by selecting YES
M73
Set low value for alarm #1; there are tw o alarms; alarm output can be set via
M77 and M78
M74
Set high value for alarm #1
M75
Set low value for alarm #2
M76
Set high value for alarm #2
M77
Set up acoustic signal
M82
Date totaliser (by day, month or year)
M85
Activate or deactivate Auto Pow er Off function (device pow ers off after 4
minutes of inactivity to save energy)
M90
View signal strength, signal quality, time ratio in the upper right corner
M91
View ratio betw een measured and calculated total transit time; if all pipe
parameters have been entered correctly and the sensors have been installed
correctly, the ratio should be around 100
% ±3 %; if not so, check all
parameters and sensor installation
M92
View estimated fluid sound velocity; if there is an apparent difference to the
actual fluid sound velocity, check all parameters and the sensor installation
M93
View total transit time and delta time (transit time difference)
M94
View Reynolds number and pipe factor the device uses
M+0
Read out 64 recorded data (Sw itch-on / sw itch-off date and time, time at w hich
flow was measured)
M+1
View total operating time of device
M+2
View date and time of last sw itch-off
M+3
View last volume flow before sw itch-off
M+4
View how often device w as sw itched on
M+5
Scientific calculator (complicated to use)
M+6
Set flow velocity
M+7
Select protocol
M+8
Waveform of the received signal (identical w ith M06)
E
n
g
lis
h