background image

2

upon the coupling type or time constant of the external
charge amplifying device. This characteristic may be
fixed or adjustable, depending on the charge amplifier.
Please refer to the appropriate charge amplifier
specification sheet for information on low-frequency
characteristics.

Figure 1.

Assorted Mounting Configurations and Their

Effects on High Frequency

4.1 STUD MOUNT

This mounting technique requires smooth, flat contact
surfaces for proper operation and is recommended for
permanent and/or secure installations. Stud mounting is
also recommended when testing at high frequencies.

Note:

Do NOT attempt mounting on curved, rough, or

uneven surfaces, as the potential for misalignment and
limited contact surface may significantly reduce the
sensor’s upper operating frequency range.

Figure 2.

Mounting Surface Preparation

STEP 1:

First, prepare a smooth, flat mounting surface,

then drill and tap a mounting hole in the center of this
area as shown in Figure 2 and in accordance with the

Installation Drawing

for the specific sensor that is

being mounted

.

A

precision-machined

mounting

surface

with

a

minimum

finish

of

63

µ

in

(0.00016

mm)

is

recommended. (If it is not possible to properly prepare
the test structure mounting surface, consider adhesive
mounting as a possible alternative.) Inspect the area,
checking that there are no burrs or other foreign
particles interfering with the contact surface.

STEP 2:

Wipe clean the mounting surface and spread

on a light film of grease, oil, or similar coupling fluid
prior to installation.

Figure 3.

Mounting Surface Lubrication

Adding

a

coupling

fluid

improves

vibration

transmissibility by filling small voids in the mounting
surface and increasing the mounting stiffness. For semi-
permanent mounting, substitute epoxy or another type of
adhesive.

STEP 3:

Screw the mounting stud into the base of the

accelerometer

and

hand-tighten.

Then,

screw the

sensor/stud assembly into the prepared tapped hole and
tighten to the recommended mounting torque as
indicated on the

Installation Drawing

.

Note:

It is important to use a torque wrench during this

step.

Under-torquing the sensor may not adequately

couple the device; over-torquing may result in stud
failure.

4.2 ADHESIVE MOUNT

Adhesive mounting is often used for temporary
installation or when the test object surface cannot be
adequately prepared for stud mounting. Adhesives like
hot glue and wax work well for temporary mounts; two-
part epoxies and quick-bonding gels provide a more
permanent mount.

Note:

Adhesively mounted sensors often exhibit a

reduction in high-frequency range.

Generally, smooth

surfaces and stiff adhesives provide the best frequency
response.

Содержание 357E91

Страница 1: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 2: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 3: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 4: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 5: ...for most models These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high temperatures preclude the use of accelerometers...

Страница 6: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 7: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 8: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 9: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 10: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 11: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 12: ...el Sealing Sensor Housing Hermetic Hermetic Sealing Connector Welded Hermetic Welded Hermetic Size Height x Length x Width 66 in x 1 16 in x 66 in 16 7 mm x 29 5 mm x 16 7 mm Weight without cable 1 8...

Страница 13: ...rs without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN A UPDATED MOUNTING HOLE DEPTH 46617 656 16 66 290 7 37 1 16 29 5 265 6 73 72 18 2 120 4 3048 102 1 28 32 5 286 THRU...

Отзывы: