Model Number

357E91

CHARGE OUTPUT ACCELEROMETER

Revision: A

ECN #: 46623

Performance

ENGLISH

SI

Sensitivity(± 10 %)

5 pC/g

.51 pC/(m/s²)

Measurement Range

± 1000 g pk

± 9800 m/s² pk

Frequency Range(± 5 %)

2.5 kHz

2.5 kHz

Frequency Range

(± 1 dB)

4 kHz

4 kHz

Resonant Frequency

 13.5 kHz

 13.5 kHz

Non-Linearity(per full scale range)

 2.5 %

 2.5 %

Transverse Sensitivity

 5 %

 5 %

Environmental

Overload Limit(Shock)

± 2000 g pk

± 19,600 m/s² pk

Temperature Range(Operating)

-67 to 1200 °F

-55 to 649 °C

Temperature Range

(Survivable Limit)

-67 to 1400 °F

-55 to 760 °C

Temperature Range

(Cable Termination)

-67 to 900 °F

-55 to 482 °C

Temperature Response

See graph

See graph

[2]

Radiation Exposure Limit(Integrated Neutron Flux)

1 E10 N/cm²

1 E10 N/cm²

Radiation Exposure Limit

(Integrated Gamma Flux)

1 E8 rad

1 E8 rad

Electrical

Capacitance

1000 pF

1000 pF

[1][2]

Insulation Resistance(@ 1200°F[649°C])

 10 kohm

 10 kohm

Insulation Resistance

(@ 70° F [21°C])

 100 MOhm

 100 MOhm

Electrical Isolation

Signal isolated from case

Signal isolated from case

Physical

Sensing Element

UHT-12™

UHT-12™

Sensing Geometry

Shear

Shear

Housing Material

Inconel

Inconel

Sealing(Sensor Housing)

Hermetic

Hermetic

Sealing

(Connector)

Welded Hermetic

Welded Hermetic

Size (Height x Length x Width)

.66 in x 1.16 in x .66 in 16.7 mm x 29.5 mm x 16.7 mm

Weight(without cable)

1.8 oz

50 gm

[2]

Electrical Connector

Integral Hardline Cable

Integral Hardline Cable

Cable Length

10 ft

3.05 m

Cable Type

MI Hardline Cable

MI Hardline Cable

Cable Termination

10-32 Coaxial Jack

10-32 Coaxial Jack

Mounting

Through Hole

Through Hole

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

NOTES:

[1] Additional cable length will add approx. 100pF/ft.
[2] Typical.

SUPPLIED ACCESSORIES: 

Model 081C108 Mounting Screw, 1/4-28x1, Inconel (1)
Model ACS-1 NIST traceable frequency response (10 Hz to upper 5% point). (1)

Entered: LK

Engineer: GJR

Sales: JC

Approved: BAM

Spec Number:

Date: 3/30/2017

Date: 3/30/2017

Date: 3/30/2017

Date: 3/30/2017

66378 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

Содержание 357E91

Страница 1: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 2: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 3: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 4: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 5: ...for most models These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high temperatures preclude the use of accelerometers...

Страница 6: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 7: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 8: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 9: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 10: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 11: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 12: ...el Sealing Sensor Housing Hermetic Hermetic Sealing Connector Welded Hermetic Welded Hermetic Size Height x Length x Width 66 in x 1 16 in x 66 in 16 7 mm x 29 5 mm x 16 7 mm Weight without cable 1 8...

Страница 13: ...rs without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN A UPDATED MOUNTING HOLE DEPTH 46617 656 16 66 290 7 37 1 16 29 5 265 6 73 72 18 2 120 4 3048 102 1 28 32 5 286 THRU...

Отзывы: