Drawing Number: 21105 
Revision: NR

 

 

 

 

OPERATION MANUAL FOR 

BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR 

FOR AMMUNITION TESTING 

Model 090B 

 

 1 

 

1.0 

INTRODUCTION 

 
The  Series  090B  Calibration  Adaptors  are  designed 
to facilitate calibration of the Model 117B Conformal 
Pressure. 
 
An  actual  empty  cartridge  case  is  pressurized  with 
precisely  known  hydraulic  pressure  and  the 
corresponding  output  of  the  sensor  is  read  at  each 
pressure point to full scale. 
 

2.0 

DESCRIPTION 

 
Refer to Drawing No. 090-2    -90 included as part of 
this manual. 
 
The  Model  090B  Adaptor  Assembly  consists  of  a 
pressure  chamber  (Item  1)  a  pressure  chamber 
adaptor  (Item  2),  a  cartridge  case  retainer  (Item  3), 
and a retaining nut (Item 4). 
 
The  pressure  chamber  adaptor  (Item  2)  contains  an 
O-ring  to  affect  a  pressure  seal  when  the  adaptor  is 
pushed  into  the  cartridge  case  mouth.    The  opposite 
end  of  this  item  is  designed  to  mount  directly  into  a 
standard  mounting  port  for  the  PCB  Models  118A, 
108A, 119A and 109A pressure sensors.  This feature 
precludes  the  need  for  high  pressure  fitting,  tubing, 
etc. thereby, keeping the total pressurized volume to a 
minimum. 
 

3.0 

INSTALLING THE SENSOR 

 
The sensor is installed in the empty pressure chamber 
(Item 1) so that the flushness of the curved diaphragm 
with the I.D. of the test chamber can be observed. 
 
Before  installing  the  sensor  into  the  Model  090B 
Adaptor,  read  the  installation  instructions  supplied 
with  the  specific  sensor  for  which  this  adaptor  is 
designed. 
 
Follow  these  instructions  carefully  paying  particular 
attention  to  the  depth  adjustment.    The  rotational 
alignment  is  automatically  controlled  by  the 
alignment device incorporated in all 117B Conformal 
Sensors. 
 

 
 

 

4.0 

INSTALLATION OF THE 
CARTRIDGE CASE 

 
(Refer to Installation Dwg.  No.  090-2    -90) 
 
It  is  important  that  cartridge  case  selected  for 
calibration  be  identical  in  all  respects  including 
material, to the lot of ammunition undergoing test.  It 
is  advisable  that  the  test  unit  be  selected  from  the 
same  lot  as  the  ammunition  under  test  for  greatest 
accuracy of test results. 
 
For small arms cartridge cases, it is necessary to use a 
case with the primer in place, but fired. 
 
Note:    Do  not  use  a  cartridge  case  with  an  unfired 

primer or a case which has drilled holes. 

 
If a sample cartridge is a loaded round, remove bullet 
and powder, then fire the primer. 
 
On  certain  calibers  an  O-ring  is  provided  in  the 
cartridge  case  retainer  (Item  3)  to  effect  a  pressure 
seal at the boltface end of the case. 
 
Inspect  the  boltface  surface  for  irregularities  or 
imperfections  in  the  area  of this O-ring,  which could 
preclude a pressure seal. 
 
Note:    For  shotshell  installation,  consult  enclosed 
drawing  for  necessary  modification  to  the  empty 
shotshell  case.    The  case  must  be  cut  to  a  specified 
overall  length  to  fit  properly  into  the  calibration 
fixture  due  to  inconsistencies  in  the  designs  of  the 
various makes of ammunition. 
 

4.1 

STEP-BY-STEP INSTALLATION 
PROCEDURE 

 
1.  Remove Items 2, 3, and 4 from chamber body 

(Item 1). 

 

Note:    For  metallic  cartridges,  it  is  extremely 
important that the mouth of the cartridge case be 
chamfered  to  allow  easy  entry  of  the  O-ring  on 
the  pressure chamber adaptor.   This operation is 
easily  performed  with  a  hand  held  90° 
countersink or  a sharp de-burring tool.  If this is 
not  done,  O-ring  life  will  be  drastically 
shortened. 

Содержание 117B44

Страница 1: ...or 30 07 Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...munition under test to take into account transmissibility characteristics of the cartridge case for each lot of ammunition Because a certain amount of pressure is required to fully obturate the cartri...

Страница 7: ...r two percent of full scale is specified for quartz sensors An acceptable instrument is then defined as one whose data points all fall within this error band The sensitivity of an instrument so calibr...

Страница 8: ...nce between the cartridge case and the chamber wall Obturation of the cartridge occurs in this region However when obturation is complete the sensor output is then linear with pressure from this point...

Страница 9: ...e data from 20 consecutive calibration runs performed on a conformal sensor yielded a slope of 04 pC psi and an intercept of 7 500 psi determined by constructing the best fitting straight line as desc...

Страница 10: ...can be further refined to include a variable voltage source as a bias in series with the output of the charge amplifier This voltage bias can be set to simulate the offset pressure and then the true p...

Страница 11: ...cale pressure to find full scale picocoulombs 40 000 psi x 250 pC psi 10 000 pC Set the charge calibrator for 10 000 pC and apply this signal to the input of the amplifier Now adjust the gain of the o...

Страница 12: ...Drawing Number 21210 Revision NR AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA 7 Adds offset pressure not available from PCB...

Страница 13: ...gement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thi...

Страница 14: ...ct a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each tim...

Страница 15: ...to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utili...

Страница 16: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 4...

Страница 17: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 5...

Страница 18: ...APTOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail in...

Страница 19: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 20: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 21: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 22: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 23: ...tment The rotational alignment is automatically controlled by the alignment device incorporated in all 117B Conformal Sensors 4 0 INSTALLATION OF THE CARTRIDGE CASE Refer to Installation Dwg No 090 2...

Страница 24: ...of torque on this nut to affect a pressure seal 4 Connect 117B to an electrostatic charge amplifier such as PCB Model 462A or 462B52 on long time constant Zero the system electrically by moving front...

Страница 25: ...for tabulating this data and calculating the standard deviation is shown in figures 1 and 2 7 0 UTILIZING THE DATA See An Improved Technique for Utilization of Conformal Ballistics Sensor Calibration...

Страница 26: ...Drawing Number 21105 Revision NR OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B 4 Figure 1...

Страница 27: ...Drawing Number 21105 Revision NR OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B 5 Figure 2...

Страница 28: ...or 30 07 Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Страница 29: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 30: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 31: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 32: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 33: ...Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 45 oz 13 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvem...

Страница 34: ...APTOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail in...

Страница 35: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 36: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 37: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 38: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 39: ...or 30 07 Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E...

Страница 40: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 41: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 42: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 43: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 44: ...1 23 HOLD THIS DIMENSION TO OBTAIN FLUSH MOUNT 1 2 3 4 5 2 1 4 REVISIONS REV DESCRIPTION DIN F UPDATED 065A19 NOTE 50698 117 2440 90 PCB Piezotronics Inc claims proprietary rights in the information d...

Отзывы: