Select Value
◆
◆
Select Value
◆
◆
Select Value
◆
◆
Enter Value
◆
◆
Select Value
◆
◆
Enter Value
◆
▼
I
P
▼
S
u
b
n
t
▲
▲
◆
Enter
Value
◆
I
P
1
◆
Enter
Value
I
P
2
5
5
From IP 1 to IP 4
◆
Enter
Value
◆
s
u
b
1
◆
Enter
Value
s
u
b
2
5
5
From Sub 1 to Sub 4
Page 18
SER - SETTING THE SERIAL COMMUNICATION PORTS
CON1
COM1:
It defines the use of the COM1 serial port.
None:
Serial communication OFF
Contn:
Continuous transmission of the weight string. It can be used,
for example, to drive a remote display. See details in the relevant
paragraph.
Deman:
When the Operator presses the front button or through Input
2, a string of weight is transmitted. The command is accepted if the
weight is stable. Between two consecutive transmissions the weight
must have a variation of at least the parameter “Weight delta”.
Autom:
It’s automatically transferred to a string of weight when the
weight stabilizes at a value higher than the minimum weight (20 divi-
sions). Between two consecutive transmissions the weight must have
a variation of at least the value set in the parameter “Delta Weight”.
Slave:
ASCII protocol. See details in the relevant paragraph.
Modbs:
MODBUS RTU (slave) protocol. See details in the relevant
paragraph.
Selectable Values:
None
Contn
Deman
Autom
Slave
Modbs
Default: Modbs
baud1
BAUD RATE COM1
It defines the RS232 serial port baudrate.
The value must be set to the same value of the PC / PLC or remote
display.
Selectable Values:
2400
4800
9600
19200
38400
57600
115200
Default: 9600