Load Weight
Enter Value
◆
With
unload
scale
Page 15
CALIB - CALIBRATION
The calibration method below, is used to correct or reduce the linearity error of the weighing system.
The calibration should be performed with the use of sample weight or pre-weighed product on a sample
weighing system.
Before proceeding with the calibration of the full scale, always perform the zero calibration.
During the calibration phase, the display shows the weight intermittently with the inscription
CAL
.
ATTENTION: at power off without exiting the set-up menu, the programming executed are not stored.
NB In the event that after calibration the system has linearity errors, verify that the structure weighed is
completely free from mechanical constraints.
ZERO CALIBRATION
Perform the operation when the system is empty (including the fixed
tare) and the weight is stable. The zero of the system is done by pres-
sing the
t
key. The display shows
2-Ok
confirming the operation.
The weight displayed resets and the display shows
CAL
alternated
by
0
. It is possible to repeat this operation more times.
CALIBRATION OF FULL SCALE
Prior F.S. calibration load the sample weight on the system and wait
for the stabilization; the display shows a weight value.
Press the
s
key to adjust the weight. The display shows
00000
with
the first digit flashing.
Use the
s
or
t
keys, enter the weight value starting with the first
digit flashing. Switch to the next digit by pressing
u
. The confir-
mation of the last digit performs the correction of the weight. The
display shows
CAL
alternated to the weight entered.
If the weight value is higher than the resolution, it is not accepted
and the display shows an error message for a few seconds.
You can always repeat the F.S. calibration.
EXIT FROM CALIBRATION MENU
The exit from the menu
CALIb
is done by pressing
u
key.
The display shows
CALIb
. To store the new setting and exit the setup
menu, press the
s
key.