DISPLAY INFO
When the instrument is switched ON the test-display is performed. In sequence there are: software
code, software version and hardware version.
profb
Profibus DP hardware module
It’s important to communicate these data in the event of a request for assistance.
ERRORS NOTIFICATION
In the operating mode, the display can report the following error codes.
-
Overload: The weight applied to the load cell exceeds by more than 9 divisions the maximum
capacity of the weighing system.
O-L
O-L : No signal from the load cells or outside of the field of measurement mV/V.
noCAL
NOCAL: Transmitter not calibrated. Recalibration needed.
Efbus
EFBUS: Fieldbus interface absent or not working.
n-coN
N-COM: Fieldbus interface of the instrument is not connected to the network.
E-CrC
E-CRC: CRC error during communication with the fieldbus interface of the instrument.
ErNEN
ERMEM: Error in E2PROM. You can reprogram the instrument to the factory settings, erasing
any calibration by pressing
u
key.
profn
Profinet hardware module