background image

Instruction Sheet

012-15719A

Carbon Dioxide ISE Probe

PS-3517

3

2

4

5

6

Internal 

Glass 

Electrode

Notch

Key Pin

Cap

Tape

Threaded 

area

Boot

1

Cover

DO NOT let 

the BNC 

connector at 

the top of 

the unit get 

wet. 

Included Equipment

Item

Included Equipment

Item

Carbon Dioxide ISE Electrode Body (2)

1

CO2 Standard Solution 1000 ppm

4

Carbon Dioxide ISE Internal Electrode

2

CO2 Reference Fill Solution

5

BNC Male-to-Female Cable

3

CO2 Ionic Strength Adjuster (ISA)

6

Содержание PS-3517

Страница 1: ...d area Boot 1 Cover DO NOT let the BNC connector at the top of the unit get wet Included Equipment Item Included Equipment Item Carbon Dioxide ISE Electrode Body 2 1 CO2 Standard Solution 1000 ppm 4 Carbon Dioxide ISE Internal Electrode 2 CO2 Reference Fill Solution 5 BNC Male to Female Cable 3 CO2 Ionic Strength Adjuster ISA 6 ...

Страница 2: ...e tip of the internal glass electrode Keep the boot for later use Note Soaking the internal glass electrode in internal filling solution or pH 4 01 buffer for at least 1 hour before proceeding with the following steps will ensure fast response 2 Remove an electrode body from the box Using the solution provided fill the electrode body with about 3 5 mL of Reference Fill Solution 3 Rinse the interna...

Страница 3: ...ctrode response follow the steps below 1 Check the slope of the internal glass electrode Remove the internal glass electrode from the electrode body Connect the internal glass electrode to the Wireless pH Sensor or to a pH mV meter Put the internal glass electrode in pH 7 00 buffer The annular ceramic junction should be immersed in the solution Record the reading in mV when it is stable Remove the...

Страница 4: ...100 mL of the sample into a 150 mL beaker Add 1 mL of ISA Stir thoroughly Check the pH value of the standard 6 Rinse electrode with Dl water blot dry and place in the sample beaker Wait for a stable reading and the sample concentration will be displayed on the mea suring device 7 Determine the sample concentration using the data analysis tools in the software Electrode Storage Short Term over nigh...

Страница 5: ...ay be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulations that vary by country and region It is your responsibility to recycle your electronic equipment per your local environmental laws...

Отзывы: