background image

®

Wireless Force-Acceleration Sensor

PS

-

3202

10

013-14710C

Appendix B: Calibration

Calibration is not necessary, especially if you are 

measuring a change in force rather than absolute force 

values. However, it is possible to calibrate the sensor. 

Prepare for Calibration

Calibration will need a 1 kilogram mass, the hook 

attachment, and a horizontally mounted support rod to 

hold the sensor. The sensor will need to be “connected” 

with a tablet or computer, and the data collection 

software (for example, SPARKvue) should be running.

Using SPARKvue Software for Calibration

See the SPARKvue Help for information about 

calibrating: 

In SPARKvue, select the HELP button (

) in any 

screen, such as the Home Screen.

1. Mount the sensor on a 

horizontal rod. Screw the hook 

attachment into the sensor. 

2. Click (or press) the Experiment 

Tools button ( ).

The Experiment Tools screen 

opens.

3. Click Calibrate Sensor.

The Calibrate Sensor: Select Measurement screen 

opens.

4. Click the Sensor box and click the sensor to be 

calibrated.

5. Click the Calibration Type box and click a calibration 

type. (For this example, click “2-point”.)

6. Click Next.

The Calibrate Sensor Enter Values screen opens.

7. Hang the 1 kilogram mass on the hook attachment. 

8. The gravitational force on the mass is pulling in the 

negative direction at -9.8 newtons (N). Under 

Calibration Point 1, click the Standard Value box 

and enter the known force value (that is, -9.8).

9. Under Calibration Point 1, click Read From Sensor.

The value measured by the sensor is transferred to 

the Sensor Value box.

10. Remove the 1 kilogram mass from the hook 

attachment. The net force now is zero N. 

11. Under Calibration Point 2, click the Standard Value 

box and enter the second known force value (i.e., 0).

12. Under Calibration Point 2, click the Read From 

Sensor box.

The second value measured by the sensor is 

transferred to the Sensor Value box.

13. Click OK.

Using PASCO Capstone for Calibration

See the PASCO Capstone Help for information about 

calibrating:

In PASCO Capstone, select PASCO Capstone Help 

from the Help menu, or press F1.

1. Mount the sensor on a horizontal rod. Screw the 

hook attachment into the sensor.

2. Click Calibration ( ) in the Tools palette.

3. Choose the measurement you would like to calibrate 

now: Force Measurement.

4. Click Next.

5. Choose the type of calibration you would like to 

perform: Two Standards (2 point).

6. Click Next.

7. Hang the 1 kilogram mass on the hook attachment. 

8. The net force on the sensor is -9.8 N. Enter the 

value in the Standard Value text box.

9. When the Current Value stabilizes, click Set Current 

Value to Standard Value.

10. Click Next.

11. Remove the 1 kilogram mass from the hook 

attachment. The second force value is zero N.

12. Enter the second known force value (i.e., 0) in the 

Standard Value text box.

13. When the Current Value stabilizes, click Set Current 

Value to Standard Value.

14. Click Next.

Содержание PS-3202

Страница 1: ...res acceleration in three dimensions X Y and Z The PASCO Data Collection Software can show the resultant acceleration The sensor uses a micro electromechanical system MEMS device The sensor is designe...

Страница 2: ...shine green when the battery is charged The battery is partially charged at the factory Initial charging time may be three hours or longer depending on the power source and the condition of the batter...

Страница 3: ...o when the force or acceleration is actually zero This is a normal behavior that is corrected when you zero or tare the sensor using the PASCO Data Collection Software In SPARKvue the process for sett...

Страница 4: ...SB Cable into the micro USB port on the end of the sensor Connect the other end of the Micro USB Cable to a USB port on the computer or into a powered USB hub connected to the computer Zero Tare the F...

Страница 5: ...ating with the computer software or tablet application try restarting the software or application If the problem remains press and hold the ON button for 10 seconds and then release Start the sensor i...

Страница 6: ...is 1 0 g If the sensor is placed label side down so that the back is facing upward the Acceleration Z axis reading is 1 0 g Mounting the Sensor Mounting the Sensor on a Rod Slide the sensor onto a rod...

Страница 7: ...eeks or more because full day continuous sampling would be unusual Even in the most extreme case with a high sample rate the Force Acceleration Sensor battery would support a full day of use before ne...

Страница 8: ...ference and 2 this device must accept any interference received including interference that may cause undesired operation CE Statement This device has been tested and found to comply with the essentia...

Страница 9: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Страница 10: ...int 1 click the Standard Value box and enter the known force value that is 9 8 9 Under Calibration Point 1 click Read From Sensor The value measured by the sensor is transferred to the Sensor Value bo...

Страница 11: ...each point is reset by associating a known standard value for instance the temperature of ice water with a raw input measurement that the sensor sends out when it is in that standard In a one point ca...

Страница 12: ...used to make one sensor agree with another sensor Due to normal variation among probes a second probe might read consistently higher than the first probe Normally this difference would be insignifica...

Отзывы: