Model No. PS-2230
Optical Dissolved Oxygen Probe Replacement Cap Calibration
13
012-14325A
SPARKvue: Calibration Code
Coefficients
1.
In SPARKvue, touch or click the “DO2
Saturation” measurement and then
touch/click “Show” to open a Graph dis-
play screen.
2.
In the Graph display, touch/click the
“Experiment Tools” icon (
) to open
the Experiment Tools menu.
3.
In the Experiment Tools menu,
touch/click “Configure Sensor” and
then touch/click “OK” to open the Sen-
sor Configuration screen.
4.
The screen shows three selections: “Edit
Sensor Properties”, “Edit Timer Proper-
ties” and “Edit Spectrometer Proper-
ties”. Touch or click Edit Sensor
Properties to show the sensor menu, and
then touch/click “Optical Dissolved
Oxygen Sensor”. The Edit Sensor Prop-
erties screen changes to show “Salinity
Constant” and “Cap Coefficient”.
5.
In the text area for “Cap Coefficient”,
enter the Calibration Code Coefficients
(K1 through K5 and KC) consecutively
without spaces. Touch or click “Set Cap
Coefficient” to save the coefficients. A
small window opens to confirm that the
procedure was successful.
6.
Touch or click “OK” to return to the Graph display screen.
Sensor
menu
Enter Calibration Code
Coefficients here