Model No. PS-2230
Optical Dissolved Oxygen Probe Replacement Cap Calibration
11
012-14325A
1.
In Capstone, click the Hardware Setup icon (
) in the Tools palette to open the
Hardware Setup panel.
2.
In the Hardware Setup panel,
click the “Properties” icon (
)
in the lower right corner to
open the Sensor Properties win-
dow.
3.
Click “Cap Coefficient” and
then enter the Calibration Code
characters (K1 to K5 and KC)
consecutively without spaces.
4.
Click “OK” to close the Proper-
ties window.
5.
Perform a one-point or a
two-point calibration. (Go to
the PASCO web site at
www.pasco.com and enter
“Optical Dissolved Oxygen
Sensor” or “PS-2196” in the
Search window. Download the
instruction manual PDF file
under the User Resources tab.)
SPARK SLS: Calibration Code Coefficients
1.
As in the calibration process, touch “Experiment Tools” in a display to open the
“Experiment Tools” screen.
2.
In the Experiment Tools screen, touch
“Configure Sensor” to open the Sensor
Configuration screen.