Advanced Water Quality Sensor Optical Dissolved Oxygen Probe Replacement Cap Cal-
12
012-14325A
3.
In the Sensor Configuration screen, touch “Edit Sensor Properties”. On the drop down
menu, touch “Optical Dissolved Oxygen Sensor”.
4.
The the “Edit Sensor Properties: Opti-
cal Dissolved Oxygen Sensor” screen
opens. Touch “Set Cap Coefficients” to
open the Enter Text screen.
5.
Use the keyboard in the Enter Text
screen to enter the Calibration Code
Coefficients (K1 to K5 and KC). Enter
the coefficients consecutively and
without spaces. After the Calibration
Code Coefficients have been entered,
touch “OK”.
•
NOTE: An error message screen
will appear if the sensor is not con-
nected to the interface, or the Cali-
bration Code Coefficients were not
entered properly. Press “OK”.
Check the sensor connection or
re-enter the Calibration Code
Coefficients.
6.
Perform a one point or two point cali-
bration with “known standard” solu-
tions. (Go to the PASCO web site at www.pasco.com and enter “Optical Dissolved
Oxygen Sensor” or “PS-2196” in the Search window. Download the instruction man-
ual PDF file under the User Resources tab.)