QUALITY ASSURANCE PLAN
The following section describes the Quality Assurance Plan recommended by the
manufacturer of the PAS Vr. Alcohol Screening & Verification System.
The manufacturer recommends the PAS Vr. System calibration be checked every six (6)
months and sensitivity be checked monthly, or more frequently if the operator feels it is
necessary because of testing schedules. In the experience of the manufacturer a
monthly check of sensitivity using a known source of alcohol is adequate to assure the
instrument is functioning under normal operating conditions (see product specifications).
Additionally, regular visual inspection of the system to confirm the intake port or exit port
is not blocked, and each light (LED) checked to be sure they display properly when
activated. Note all inspections and testing in the logbook by serial number. In the event
of an observed malfunction or failure of the PAS Vr. to test accurately call
1-800-660-SNIF for instructions or guidance.
Compliance with the following procedures will assure reliable testing, calibration, and
operation of the PAS Vr. System.
1. Minimum Intervals for Testing System Performance
2. Tolerance (See Product Specifications)
3. Inspection, maintenance and calibration
4. Record Keeping - Logs