PAS Systems International PAS V Скачать руководство пользователя страница 7

 

 

Readying the PAS V

 

 

 1.     

Be  sure  the  battery  is  installed  and  fully  charged,  with  the  flashlight  ON.  There 
should be no blinking red battery light. 

 

2.      Inspect the inlet/outlet ports of the PAS V to make sure that they are not clogged. 

 

3.      Conduct an air blank:     

 

 

a.  Turn  on  the  flashlight  and  take  an  air  sample  by  pressing  the  Sensor  Control 

Button; the green indicator light comes on  

 

b.  Observe  the  light  bar  for  15-20  seconds.  If  only  the  small  green  sampling 

indicator light comes on, the air blank check is OK. 

 

4.      Clear a fuel cell overload (if one or more of the bar-graph indicator  
           lights comes on during air blank):  

 

a.  Turn the flashlight on to activate fuel cell heater (check that the HTR light is on). 

This speeds the cell’s response, clearance and recovery time. 

 

b.  Turn off the PAS V. This also speeds the recovery of the fuel cell. 

 

c.  After 5 minutes, repeat the air blank making sure that the flashlight is on. If none 

of the bar-graph indicator lights come on (after observing for 15-20 seconds), the 
air blank check is OK. 

 

d.  If  however,  one  or  more  of  the  indicator  lights  come  on  again,  this  clearance 

process should  be repeated.  If after two attempts  to clear the unit, the indicator 
lights  are  still  coming  on  when  the  air  blank  is  performed,  there  are  two 
possibilities: 

 

1.  The  unit  requires  an  internal  zero  adjustment.  To  have  this  adjustment 

made, the unit has to be returned to the manufacturer. 

 

2.  The  air  in  the  test  room  contains  a  significant  amount  of  alcohol.  To 

determine if this is the case, air blanks should be made away from the test 
room to see whether the unit gives a zero reading when removed from the 
test area. 

 
5.      Once the air blank has been performed with a satisfactory result, the PASV unit 

is ready for sensitivity testing.   

NOTE

:

 Be sure the room air is free from alcohol and that the ambient temperature 

is above 50

°

F. Do not have jars of stock solution or other alcohol open in the room 

during testing.     

 

NOTE:  Throughout  the  procedure,  the  test  conductor  should  check  for  a  flashing 
battery  light  to  ensure  that  the  battery  is  not  running  down  during  the  testing 
procedure. 

Содержание PAS V

Страница 1: ...Passive Alcohol Sensor CALIBRATION MANUAL PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com Copyright 2015 by PAS Systems International Inc All Rights Rese...

Страница 2: ......

Страница 3: ...Bath Solution optional 2 Operating the Alcohol Wet Bath Simulator 3 Readying the PAS V 5 PAS V Sensitivity Checking 6 PAS V Calibration 7 Dry Gas Method Equipment Checklist for Dry Gas Sensitivity Che...

Страница 4: ...ece that connects the outlet of the wet bath simulator to the intake port of the PAS V Available from PAS Systems A NHTSA approved Wet Bath Simulator 500 milliliters of 005 ethanol alcohol solution To...

Страница 5: ...volumetric flask and fill the flask up with distilled water This completes your mixture of stock solution You may keep your stock solution in a refrigerator for mixing calibration solution as necessar...

Страница 6: ...jar from the simulator by unscrewing it from the top assembly Caution DO NOT remove Top Assembly and expose the heating element to open air with power turned ON This will result in damage to the heati...

Страница 7: ...hts come on after observing for 15 20 seconds the air blank check is OK d If however one or more of the indicator lights come on again this clearance process should be repeated If after two attempts t...

Страница 8: ...sampling began 6 Disconnect the Calibration Assembly and observe the bar graph display All but one of the bars should light up There should be two green four yellow and two of the three red bars lit 1...

Страница 9: ...activate the Sensor Control Button on the PAS V 6 Look for the green sensor light and the yellow pump PMP light and listen for the sound of the air pump that should come on immediately after the Sens...

Страница 10: ...ct the inlet outlet ports of the PAS V to make sure that they are not clogged 3 Conduct an air blank a Turn on the flashlight and wait a minimum of 1 2 minutes for initialization process to complete b...

Страница 11: ...r blanks should be made away from the test room to see whether the unit gives a zero reading when removed from the test area 5 Once the air blank has been performed with a satisfactory result the PASV...

Страница 12: ...d light up There should be two green four yellow and two of the three red bars lit 15 20 seconds from the time the pump stops 9 Record the number of red bars that light up this should be two Repeat th...

Страница 13: ...ough the opening and into the calibration adjustment pot screw Turn the calibration adjustment all the way up counter clockwise The adjustment is free to rotate approximately 270 Take care not to use...

Страница 14: ...Problem Reported Checked By Action Taken Date Returned to Manufacturer Date Returned to Field The example shown provides a means for recording any reports from Testers of problems with the PAS V units...

Страница 15: ......

Страница 16: ...PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Printed 10 15...

Отзывы: