Operating the Alcohol Wet Bath Simulator 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 
 

The Alcohol Wet Bath Simulator 

The  lower  (jar)  portion  is  filled  with  a  heated  solution  of  water  and  alcohol,  which 
simulates  the  blood  running  through  the  arteries  in  the  lungs.  The  upper  portion  (the 
head space) is filled with a mixture of air and alcohol that simulates the air in the lungs 
that is in contact with the blood across the cell membranes.  Air flowing out of the outlet 
pipe simulates the expired air of a person who has been drinking.     
 
In order to work properly, the temperature  of the water/alcohol solution must  be 34

°

 C 

(body temperature) and the mixture of alcohol and water must be accurate for the BrAC 
to be used in the test (.01 or .005 BrAC).   

 
Preparing Wet Bath Simulator 

 

1.      Remove the glass jar from the simulator by unscrewing it from the top assembly.     
 

Caution

: DO NOT remove Top Assembly and expose the heating element to open 

air with power turned ON.  This will result in damage to the heating element.     

 
2.      Place 500 ml of the .005 calibration solution into the simulator.     
 
3.      Assemble by tightly turning the jar back into the top assembly.     
 
4.      Add tubing to the inlet and outlet of the simulator.     
 
5.      Check the seal of  the lid on the jar by blowing into the inlet while holding outlet 

closed.  There should be resistance to the airflow.     

 
6.      Plug  in  the  simulator  and  turn  the  power  ON  and  allow  the  solution  to  heat  to 

34◦C  (approx  15  min).    The  heater  lamp  is  lit  when  the  heating

 

element  is 

heating.    The  heater  lamp  goes  off  when  the  simulator  fluid  has  reached 

34◦C.  

However, it will come back on from time to time to

 

keep the fluid at 34◦C.  When 

34°C is reached, the simulator is ready to be used.    

Outlet to PAS V 

On/Off Switch 

Heater ON Light 

Inlet Tube 

Содержание PAS V

Страница 1: ...Passive Alcohol Sensor CALIBRATION MANUAL PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com Copyright 2015 by PAS Systems International Inc All Rights Rese...

Страница 2: ......

Страница 3: ...Bath Solution optional 2 Operating the Alcohol Wet Bath Simulator 3 Readying the PAS V 5 PAS V Sensitivity Checking 6 PAS V Calibration 7 Dry Gas Method Equipment Checklist for Dry Gas Sensitivity Che...

Страница 4: ...ece that connects the outlet of the wet bath simulator to the intake port of the PAS V Available from PAS Systems A NHTSA approved Wet Bath Simulator 500 milliliters of 005 ethanol alcohol solution To...

Страница 5: ...volumetric flask and fill the flask up with distilled water This completes your mixture of stock solution You may keep your stock solution in a refrigerator for mixing calibration solution as necessar...

Страница 6: ...jar from the simulator by unscrewing it from the top assembly Caution DO NOT remove Top Assembly and expose the heating element to open air with power turned ON This will result in damage to the heati...

Страница 7: ...hts come on after observing for 15 20 seconds the air blank check is OK d If however one or more of the indicator lights come on again this clearance process should be repeated If after two attempts t...

Страница 8: ...sampling began 6 Disconnect the Calibration Assembly and observe the bar graph display All but one of the bars should light up There should be two green four yellow and two of the three red bars lit 1...

Страница 9: ...activate the Sensor Control Button on the PAS V 6 Look for the green sensor light and the yellow pump PMP light and listen for the sound of the air pump that should come on immediately after the Sens...

Страница 10: ...ct the inlet outlet ports of the PAS V to make sure that they are not clogged 3 Conduct an air blank a Turn on the flashlight and wait a minimum of 1 2 minutes for initialization process to complete b...

Страница 11: ...r blanks should be made away from the test room to see whether the unit gives a zero reading when removed from the test area 5 Once the air blank has been performed with a satisfactory result the PASV...

Страница 12: ...d light up There should be two green four yellow and two of the three red bars lit 15 20 seconds from the time the pump stops 9 Record the number of red bars that light up this should be two Repeat th...

Страница 13: ...ough the opening and into the calibration adjustment pot screw Turn the calibration adjustment all the way up counter clockwise The adjustment is free to rotate approximately 270 Take care not to use...

Страница 14: ...Problem Reported Checked By Action Taken Date Returned to Manufacturer Date Returned to Field The example shown provides a means for recording any reports from Testers of problems with the PAS V units...

Страница 15: ......

Страница 16: ...PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Printed 10 15...

Отзывы: