PAS V Calibration 

 

This procedure follows the same steps as the sensitivity check procedure except that 
after the sample has been blown into the unit, the signal gain is adjusted to the point 
where two red bars light up. 

 

NOTE

 

Monitor the red low battery light of the PAS V to ensure that it does not 

start flashing during the calibration procedure. If it does, the battery must be 
charged before proceeding.

  

 

1.  Remove the plastic plug on the back of the flashlight tube by carefully prying out 

with the small screwdriver. Inside you will see the PAS V calibration adjustment 
pot (screw). 

 

2.  Repeat Steps 1-7 for PAS V Sensitivity Checking.  

 

3. 

Pass  a  small  flat  blade  jeweler’s  screwdriver  through  the  opening  and  into  the 
calibration adjustment pot (screw). Turn the calibration adjustment all the way up 
(counter  clockwise).  The  adjustment  is  free  to  rotate  approximately  270

°

.  Take 

care not to use any force. With the calibration adjustment at maximum, all three 
red bars light up. 

 

4.  When  all  bars  of  the  display  light  up  (10-20  seconds  after  sampling),  turn  the 

calibration adjustment  down  (clockwise)  until  the top  red bar  just  goes  out.  You 
may need to make several small, soft and easy adjustment as the fuel cell climbs 
to  its  peak  reading.  If  your  adjustment  is  slightly  too  high,  you  should  see  the 
third  red  bar  flash  on  and  off.  Turn  counter  clockwise  just  enough  so  that  the 
flashing stops. Calibration is now complete.  

 

5.  Replace the plastic plug by pressing firmly. 

 

6.  Turn off the flashlight and the PAS system. The unit is now ready for operational 

use once it has been fully charged.  
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание PAS V

Страница 1: ...Passive Alcohol Sensor CALIBRATION MANUAL PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com Copyright 2015 by PAS Systems International Inc All Rights Rese...

Страница 2: ......

Страница 3: ...Bath Solution optional 2 Operating the Alcohol Wet Bath Simulator 3 Readying the PAS V 5 PAS V Sensitivity Checking 6 PAS V Calibration 7 Dry Gas Method Equipment Checklist for Dry Gas Sensitivity Che...

Страница 4: ...ece that connects the outlet of the wet bath simulator to the intake port of the PAS V Available from PAS Systems A NHTSA approved Wet Bath Simulator 500 milliliters of 005 ethanol alcohol solution To...

Страница 5: ...volumetric flask and fill the flask up with distilled water This completes your mixture of stock solution You may keep your stock solution in a refrigerator for mixing calibration solution as necessar...

Страница 6: ...jar from the simulator by unscrewing it from the top assembly Caution DO NOT remove Top Assembly and expose the heating element to open air with power turned ON This will result in damage to the heati...

Страница 7: ...hts come on after observing for 15 20 seconds the air blank check is OK d If however one or more of the indicator lights come on again this clearance process should be repeated If after two attempts t...

Страница 8: ...sampling began 6 Disconnect the Calibration Assembly and observe the bar graph display All but one of the bars should light up There should be two green four yellow and two of the three red bars lit 1...

Страница 9: ...activate the Sensor Control Button on the PAS V 6 Look for the green sensor light and the yellow pump PMP light and listen for the sound of the air pump that should come on immediately after the Sens...

Страница 10: ...ct the inlet outlet ports of the PAS V to make sure that they are not clogged 3 Conduct an air blank a Turn on the flashlight and wait a minimum of 1 2 minutes for initialization process to complete b...

Страница 11: ...r blanks should be made away from the test room to see whether the unit gives a zero reading when removed from the test area 5 Once the air blank has been performed with a satisfactory result the PASV...

Страница 12: ...d light up There should be two green four yellow and two of the three red bars lit 15 20 seconds from the time the pump stops 9 Record the number of red bars that light up this should be two Repeat th...

Страница 13: ...ough the opening and into the calibration adjustment pot screw Turn the calibration adjustment all the way up counter clockwise The adjustment is free to rotate approximately 270 Take care not to use...

Страница 14: ...Problem Reported Checked By Action Taken Date Returned to Manufacturer Date Returned to Field The example shown provides a means for recording any reports from Testers of problems with the PAS V units...

Страница 15: ......

Страница 16: ...PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Printed 10 15...

Отзывы: