Page 6-16
Lasair II Particle Counter Operations Manual
Chapter 6: Operation
Statistics Mode Sample Print
When the Lasair II particle counter is operated in statistics mode, sample printouts
have the room name appended to the sample strip. The room is printed directly
above the location name.
Silencing Alarms
If the alarm is activated, the
Main Display Screen
(and
Analog Data Display
Screen
) will blink red and yellow. If the audio alarm has been set, the unit will beep.
To silence an alarm:
Press the
Enter
key or one of the arrow buttons on the control panel until all alarms
are off. The value in alarm will remain in red on the
Main Display Screen
if it is a
particle count value (or on the
Analog Data Display Screen
, if it is an analog value).
12
||10.0|| 16084|| 16084||
The 10 micron channel:
|| total counts of 10 microns || total counts
≥
10 microns
13
Avg || Min || Max
Column headers for Environmental Data.
Average, Minimum, and Maximum are taken
over the Sample period.
14
T DEG C
T: temperature DEG C: degrees centigrade
15
25.1 || 24.94 || 25.23
Average || minimum || and maximum
temperature in degrees centigrade. See lines
12 and 13.
16
rh PCT
rh: relative humidity PCT: percent
17
15.4 || 15.33 || 15.43
Average || minimum || and maximum relative
humidity in percent. (See lines 12 and 14.)
18
Alarm: Pump
This line prints when the particle counter
senses that the flow rate is outside the
boundary of 1 CFM ±5%. (Also see line 3b.)
19
Alarm: Laser
This line prints when the particle counter
senses that the laser power is out of the
specified range. (Also see line 3b.)
Table 6-1: Printout Legend (Continued)
Содержание LASAIR II 350L
Страница 1: ...Operations Manual P N 1000000982 LASAIR II 310A 310B 510A 350L 550L Particle Counter...
Страница 13: ...Table of Contents xii This page is intentionally left blank...
Страница 17: ...List of Tables xvi...
Страница 143: ...Page 8 26 Lasair II Particle Counter Operations Manual Chapter 8 Statistics Mode This page is intentionally left blank...
Страница 155: ...Page 9 12 Lasair II Particle Counter Operations Manual Chapter 9 Recipes This page is intentionally left blank...
Страница 161: ...Page 10 6 Lasair II Particle Counter Operations Manual Chapter 10 Maintenance This page is intentionally left blank...
Страница 211: ...Appendix I USB Host Port Page I 4 Lasair II Particle Counter Operations Manual This page is intentionally left blank...