Viewing Sample Data
Lasair II Particle Counter Operations Manual
Page 6-7
This selection determines the units displayed for the Sample Volume and sample
readings.
Display: [
Δ
or
Σ
]
Determine whether data is displayed in Differential (
Δ
) or
Cumulative (
Σ
) values. This setting is saved in battery-powered RAM when the
Lasair II particle counter is shut down.
Data: [N or N/ft.
3
]
Select
N
to display raw counts. Select
N/ft.
3
to display
normalized counts, that is, counts per unit of volume.
NOTE:
If you change the sampling setup after setting your units
choice in the
Data Display Setup Screen
, the units will be over-ridden
by the display units you have chosen in the
Data Display Setup
Screen
.
Sort By: [Date and time
or Location
]
Select Time to sort by the date
and time the sample data was created. Select Location to sort by the location where
the sample was collected. Before selecting “sort by location”, location names must
have already been created and assigned to the sample.
If location is selected, the data will first be sorted first by location and then by date/
time.
This setting is saved in battery-powered RAM when the particle counter is shut
down.
Order: [A...Z or Z...A
]
Select
(A...Z)
to view data in ascending order or select
(Z...A)
to view data in descending order.
This setting is saved in battery-powered RAM when the particle counter is shut
down.
Controlling what Data is Displayed
To determine which data are displayed:
1
If the
Data Table Screen
is not open, navigate to it.
Or
2
If the
Data Table Screen
is already open but it is in the navigation mode, disable
the edit mode by pressing the
Return Arrow
icon softkey.
3
Open the
Data Filter Screen
by pressing the
Data Display
icon softkey.
The
Data Filter-Time Screen
will open.
Содержание LASAIR II 350L
Страница 1: ...Operations Manual P N 1000000982 LASAIR II 310A 310B 510A 350L 550L Particle Counter...
Страница 13: ...Table of Contents xii This page is intentionally left blank...
Страница 17: ...List of Tables xvi...
Страница 143: ...Page 8 26 Lasair II Particle Counter Operations Manual Chapter 8 Statistics Mode This page is intentionally left blank...
Страница 155: ...Page 9 12 Lasair II Particle Counter Operations Manual Chapter 9 Recipes This page is intentionally left blank...
Страница 161: ...Page 10 6 Lasair II Particle Counter Operations Manual Chapter 10 Maintenance This page is intentionally left blank...
Страница 211: ...Appendix I USB Host Port Page I 4 Lasair II Particle Counter Operations Manual This page is intentionally left blank...