Page 10-2
Lasair II Particle Counter Operations Manual
Chapter 10: Maintenance
•
DECON-QUAT 100, a quaternary ammonium solution consisting of Alkyl
Dimethyl Benzyl Ammonium Chloride and Alkyl Dimethyl Ethylbenzyl
Ammonium Chloride
•
DECON-Clean
,
consisting of detergent, 2-Butoxyethanol, Potassium
Hydroxide, and Tetrasodium Pyrophosphate
•
A 50–90% solution of isopropyl alcohol in deionized water. This solution is
useful for routine cleaning to remove smudges and hand prints.
•
A 30% ammonia solution. This solution is also useful for routine cleaning to
remove smudges and hand prints.
•
Formaldehyde in a solution of less than 50 percent
•
Ethyl alcohol in a solution of less than 50 percent
Call Instrument Service and Support at Particle Measuring Systems (1-800-557-
6363) for advice about other cleaning chemicals.
Cleaning Method
The Lasair II particle counter case is made of Kydex, a rugged plastic in the PVC
family. The polyester display screen is sealed around its edges and all case openings
are sealed with gaskets. Both Kydex and polyester are resistant to most cleaning
chemicals. Never wipe the display screen with anything that is hard or sharp.
Although the Lasair II particle counter is resistant to liquids, it is not liquid proof.
Do not submerge the particle counter in any liquid or apply large amounts of liquid.
Always wipe away free standing liquid and wet wipe the particle counter after
applying a cleaner. Never put liquids into the air intake port.
To clean the Lasair II particle counter:
1
Turn the particle counter off, unplug the power cord, and remove the battery.
2
Disconnect all other cables.
3
Remove the sample horn and place the shipping cap on the air intake tube.
OR
If you prefer to clean the particle counter and the isokinetic probe with the probe
attached to the particle counter, place the shipping cap on the intake of the isokinetic
probe.
WARNING
Always wear eye protection when working with cleaning chemicals.
Содержание LASAIR II 350L
Страница 1: ...Operations Manual P N 1000000982 LASAIR II 310A 310B 510A 350L 550L Particle Counter...
Страница 13: ...Table of Contents xii This page is intentionally left blank...
Страница 17: ...List of Tables xvi...
Страница 143: ...Page 8 26 Lasair II Particle Counter Operations Manual Chapter 8 Statistics Mode This page is intentionally left blank...
Страница 155: ...Page 9 12 Lasair II Particle Counter Operations Manual Chapter 9 Recipes This page is intentionally left blank...
Страница 161: ...Page 10 6 Lasair II Particle Counter Operations Manual Chapter 10 Maintenance This page is intentionally left blank...
Страница 211: ...Appendix I USB Host Port Page I 4 Lasair II Particle Counter Operations Manual This page is intentionally left blank...