1. CyFlow
®
Cube System
Size
Dimensions: 500 mm L x 470 mm D x 355 mm H ( 670 mm with
open display )
Weight
appr.40 kg
Maximum sound
< 70 dBA
Power level
Installation/overvoltage category 2/II
Degree of protection
IP 20
Operating Environment
Temperature:15-30
°
C
Humidity: 20-85% relative (non-condensing)
Room:
Clean environment. Direct sun light should be avoided.
Applications
Immunophenotyping, DNA Analysis, Ploidy Analysis, Apoptosis,
Microbiology, Industrial applications, 3 to 6 Colour Analysis. True
volumetric absolute counts = counting per volume
True Volumetric
Based on precise counting and mechanical fluid volume measurement
Absolute Counting
No need for reference sample or beads
Instrument Check
Partec Count Check Beads
Partec Calibration Beads 1 µm and 3 µm
Partec DNA Control PI
Set-up Time
Max. 5 minutes
Parameters
Up to 8 optical parameters: FSC, SSC, FL1, FL2, FL3, FL4, FL5, FL6
Particle Size Range
0.1 µm - 50 µm (standard cuvette)
Maximum Acquisition Speed
25,000 events/sec
Acquisition Stop Time
Event- or volume-based
Trigger
On all parameters, on multiple parameters or on single trigger
parameter, selectable in software
Data Resolution
65,536 channels (16 bit)
Service
1 - 3 years service contracts
Warranty
12 months on all parts except filters, mirrors, other quartz or glass
parts, disposables and cuvettes