![Partec CyFlow Cube Series Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/partec/cyflow-cube-series/cyflow-cube-series_instrument-operating-manual_3799403025.webp)
CyFlow
®
Cube Instrument Operating Manual
Rev018_2012-01-27
25/36
Create a vertical histogram splitter
To create a vertical histogram splitters please
move the cursor into the histogram and press the
right mouse button. Select “Built vertical splitter
region” to divide the histogram into two sections
(VS1-1 and VS1-2). The intersection can be
modified by moving the cursor into the plot and
keeping the left mouse button pressed.
Create a vertical dotplot splitter
To create a vertical dotplot splitters move the
cursor into the dotplot and press the right mouse
button. Select “Built vertical splitter region” to
divide the dotplot into two sections (VS1-1 and
VS1-2). The intersection can be modified by
moving the cursor into the dotplot and keeping the
left mouse button pressed.
Create a horizontal dotplot splitter
To create a horizontal dotplot splitters move the
cursor into the dotplot and press the right mouse
button. Select “Built horizontal splitter region”
to divide the dotplot into two sections (HS1-1 and
HS1-2). The intersection can be modified by
moving the cursor into the dotplot and keeping the
left mouse button pressed.
Create a quadrant assembly
To create a quadrant assembly within a dotplot
move the cursor into the dotplot and press the
right mouse button. Select “Built cross region”
to divide the dotplot into four sections (CR1-1,
CR1-2, CR1-3 and CR1-4). To change the
quadrants move the cursor into the dotplot and
keep the left mouse button pressed. To create
asymmetric quadrants please approach individual
points (at the border of the dotplot or at the
intersection point of the quadrants) with the left
mouse button and keep the left mouse button
pressed.
Change layout for regions
In the register chart file select “Regions”.
Individual regions can be selected with the <>
arrows. Regions can be transferred to other plots
by the function “Home plot name” or modified in
their color with “Color RGB”. To activate changes
press “Accept”.