CyFlow
®
Cube Instrument Operating Manual
Rev018_2012-01-27
11/36
Instrument starting procedure
Sheath fluid level control and/or refill
Switching on the CyFlow
®
Cube
Main
power
switch
Waste and
sheath
fluid
On/Standby
Casual/medium expertise user:
Once the Cube started no further steps are
necessary. You can directly start your
measurements!
The power supply switch is found at the back of
the Cube next to the main supply cable. The Cube
is, in default, set in a stand-by mode. The full
activation of the Cube requires pressing lightly the
on/off tactile button on the top of the machine.
The display screen must be first lift up to access
it.
This
will
start
the
embedded
computer,
automatically start the CyView
™
software and load
the last employed configuration.
Before
switching
on
the
machine,
it
is
recommended to check the levels of the sheath
fluid and waste bottles. They can be found at the
back-left
of
the
apparatus
in
a
sliding
compartment.
Make sure SHEATH bottle is filled with 800 ml of
clean, filtered, and degassed sheath fluid and is
closed with the screw top. In order to guarantee
highest quality of the measurements we highly
recommend use Partec Sheath Fluid (order no.
04-4007).
It is recommended to replace the sheath fluid at
least once a week or before any daily use.
When filling up the sheath fluid bottle make sure
no air bubbles are trapped in the yellow filter unit
inside the bottle!
Make sure WASTE bottle is empty and the screw
top is tightly closed.
The waste bottle must be emptied after and
before each user session. When using bio-
hazardeous samples, a volume of 50 ml of
hypochlorite 0.5% (Order No 04-4012) should be
introduce into the empty waste bottle for initial
disinfection.
Rear side of CyFlow
®
Cube showing the main
power switch and the sheath fluid and waste
bottle compartment.