![Partec CyFlow Cube Series Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/partec/cyflow-cube-series/cyflow-cube-series_instrument-operating-manual_3799403021.webp)
CyFlow
®
Cube Instrument Operating Manual
Rev018_2012-01-27
21/36
Cleaning Procedures
Cleaning procedure during Cube running
The Cube can be cleaned between sets of samples using the cleaning and rinsing solution (green solution,
Order No 04-4009). This procedures will allow you to reduce significantly the cross contamination and
reduce the background.
The cleaning procedure should be set as follows: connect a sample tube with distilled water to the sample
port of the cube and press START three times. The cleaning tube will be entirely aspired by the instrument
and used for cleaning.
Cleaning procedure for switching off
Before switching off the Cube, it is necessary to run the decontamination solution (violet solution, Order No
04-4010) using the clean function (Menu -> stack -> clean). Then, the same operation must be repeated
with distilled water. The Cube can be put in standby by closing CyView
™
(Menu -> Exit) and closing
WindowsXP (start-> Exit-> Switch Computer Off). To power off completely the Cube, use the main power
switch at the back panel.
The waste bottle content must be discarded accordingly to the relevant biohazard regulations.
A regular thorough cleaning of the sheath fluid bottle and exchange of the yellow filter will keep the
background in the measurements to a minimum level.
No cross contamination between samples
To minimize the carry-over of one sample to the next, a cleaning cycle between sample can be done.
Start your measurement.
Click Pause, remove you sample from the sample port, introduce a sample tube filled with distilled water
Click twice on play, the sample port will be cleaned and the distilled water removed from the tube. Your
Cube is now ready for your next measurement with no cross contamination from the previous sample.
+