icountBSplus Bottle Sampler Manual
29
USA
5
6
Sample bottles may vary in size, so the drip tray
has a ‘target’ to indicate that the bottle should be
located centrally around the drip tube.
The bottle should look like this.
7
8
Bring door front to latch
Push door fully (click); the door should sit flush.
You are now ready to test.
Содержание icountBSplus
Страница 7: ...icountBSplus Bottle Sampler Manual 7 USA EC declaration of conformity...
Страница 37: ...icountBSplus Bottle Sampler Manual 37 USA bp Results screen Browse all your tests...
Страница 53: ...icountBSplus Bottle Sampler Manual 53 USA Y Y Y Y N N N N Diptube does not stay down...
Страница 54: ...Parker Hannifin Manufacturing UK Ltd 54 USA Y Y Y Y N N N N Diptube stays down...
Страница 55: ...icountBSplus Bottle Sampler Manual 55 USA Y N Y N Diptube operation not smooth Diptube stuck...
Страница 57: ...icountBSplus Bottle Sampler Manual 57 USA icountBSplus not switching on power fault using battery Y Y N N...
Страница 58: ...Parker Hannifin Manufacturing UK Ltd 58 USA icountBSplus gives suspect counts Y Y Y Y N N N N...
Страница 59: ...icountBSplus Bottle Sampler Manual 59 USA Y Y Y N N N USB memory device not found upgrade failure...
Страница 60: ...Parker Hannifin Manufacturing UK Ltd 60 USA Y Y N N Touchscreen failure...