icountBSplus Bottle Sampler Manual
49
USA
Fitting a replacement door seal
Parts required
ACC6NW016 – Replacement door
seal pack (x5)
Procedure
1
2
Open door and remove door seal. Use
isopropanol to clean the inside of the channel
and the inside of the door, removing all traces
of silicone grease.
Using a syringe, apply silicone grease
compound to the base of the channel.
DO NOT apply this to the side walls of the
channel. Use a brush to apply this evenly
over the entire base of the channel.