icountBSplus Bottle Sampler Manual
27
USA
Sample handling and preparation
BOTTLE CLEANLINESS
Bottles should have sealing screw caps, with both parts cleaned to a suitable level in accordance
with ISO3722.
Standard Parker Hannifin bottles (ACC6NW001 (2-pack) and ACC6NW002 (50-pack)) are supplied
clean to ISO 13/11 or better in a Class 10,000 Clean Room. The bottle should remain capped until
the time of sample filling and be re-capped immediately afterwards.
SAMPLE MIXING
Sedimentation of contaminant in a sample will occur, the rate of which is dependent upon both the
fluid and particle characteristics.
Visually check the fluid sample for any large particles before testing by holding up to a light source. If
you can see large contaminants, you may want to reconsider if the sample is suitable for testing by
the icountBSplus.
Other methods of sample agitation have not been provided, as they are likely to inconsistently
distort the analysis of results. Where facilities are available, mixing can be achieved using ‘paint
shakers’ and/or an ultrasonic bath. Take care when using ultrasonic baths to avoid distortion of the
result by prolonged use, which could cause the breakdown of contaminants.
Bottle samples can be sufficiently stirred by swirling and tumbling by hand, end-over-end. Samples
should be analyzed, without delay, once agitated.
NOTE: Air bubbles are counted as contamination and may affect test results. High
viscosity fluids may take longer to degas. For further information, please contact Parker
Hannifin Manufacturing (UK) Ltd.
RESULTS
The first result from a bottle sample should be disregarded, as it could be distorted by fluid from a
previous sample.
Samples from different parts of a system will give different results. Consideration should be given to
what monitoring is desired and where to extract samples from for trend monitoring to be performed.
Ensure that there is a minimum of 30ml of sample fluid left in the bottle at the end of the test
to avoid any air being drawn into the icountBSplus system. If this minimum level has not been
maintained, please flush through with a new sample of at least 50ml.
NOTE: If there is a spillage of test fluids in the chamber, please clean the icountBSplus
with a clean lint-free cloth, with any appropriate cleaning fluids being applied to the cloth.
It is important that whatever practices you adopt, you must perform them consistently.
Содержание icountBSplus
Страница 7: ...icountBSplus Bottle Sampler Manual 7 USA EC declaration of conformity...
Страница 37: ...icountBSplus Bottle Sampler Manual 37 USA bp Results screen Browse all your tests...
Страница 53: ...icountBSplus Bottle Sampler Manual 53 USA Y Y Y Y N N N N Diptube does not stay down...
Страница 54: ...Parker Hannifin Manufacturing UK Ltd 54 USA Y Y Y Y N N N N Diptube stays down...
Страница 55: ...icountBSplus Bottle Sampler Manual 55 USA Y N Y N Diptube operation not smooth Diptube stuck...
Страница 57: ...icountBSplus Bottle Sampler Manual 57 USA icountBSplus not switching on power fault using battery Y Y N N...
Страница 58: ...Parker Hannifin Manufacturing UK Ltd 58 USA icountBSplus gives suspect counts Y Y Y Y N N N N...
Страница 59: ...icountBSplus Bottle Sampler Manual 59 USA Y Y Y N N N USB memory device not found upgrade failure...
Страница 60: ...Parker Hannifin Manufacturing UK Ltd 60 USA Y Y N N Touchscreen failure...