Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
22
1-800-343-4048
No pressure
No or Low Flow
Low Output Pressure
Low Nitrogen Purity
Change in Dewpoint
• Compressor is not ON
• Regulator is closed
• System leaks at fitting
connections
• Flow demand greater than
rated capacity
• Flow demand greater than
30 LPM
• Nitrogen Module
Malfunction
• High relative humidity
• Water carryover due to
water separator float drain
malfunction
• Water carryover due to filter
float drain malfunction
• Spent carbon modules
• Internal leak
• Water carryover due to
water separator float drain
malfunction
• Water carryover due to filter
float drain malfunction
• High relative humidity
• Total flow demand greater
than 68 SPLM for the Gas
1/ Gas 2 and source gas
outlets
• Turn on compressor
• Open pressure regulator
• Check for leaks at
connections
• Reduce flow demand
• Reduce flow demand
• Replace module
• Move to area where RH is less
than 43% at 85°F – see table
• Replace water separator float
drain
• Replace filter float drain
• Replace carbon modules
• Check for leaks
• Replace water separator float
drain
• Replace filter float drain
• Move to area where RH is less
than 43% at 86°F – see table
• Reduce flow demand
NitroFlow TG2 Generator Troubleshooting
Symptom
Probable Cause
Corrective Action
• Internal leak
• External leak
• Compressor not at pressure
• Particulate filter plugged
• Membrane rupture
• Check for leaks
• Check for leaks
• See compressor section
• Replace filter
• Replace membrane