Bulletin TI-72730J 

Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

7

1-800-343-4048

Problem 

Possible Cause

No display/reading 

Sensor wires not connected;  check wire strip length, polarity.

Display varies 

Check process;  check sensor light;  check sample lines for  

 

 leaks

Alarm stays on 

Check set points

Limited range during calibration

 

Replace sensor (P/N 72695)

Troubleshooting

Don’t Forget To:

Serial Numbers

1

  Complete and mail or fax in your warranty registration card

2

  Keep your product certification in a safe place.

3

  Call the Technical Services Department at 

800-343-4048, 

8AM to 5PM Eastern Time  

 

with any questions (North America only). For other locations, please contact your local  

 representative.

The serial number for the unit is located on the bottom panel.  For your own records, and in 
case service is required, please record the following:

DATE IN SERVICE 

 

 

    SERIAL NO.

Please have the serial number available when calling for assistance.

WARRANTY (NORTH AMERICA ONLY)

(FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE)

Parker Hannifin guarantees to the original purchaser of this product, that if the product fails or is defective 
within 12 months from the date of purchase, when this product is operated and maintained according 
to the instructions provided with the product, then Parker guarantees, at Parker’s option, to replace the 
product, repair the product, or refund the original price for the product.  This warranty applies only to 
defects in material or workmanship and does not cover: ring and valve wear on compressors, routine 
maintenance recommended by the instructions provided with this product, or filter cartridges.  Any 
modification of the product without written approval from Parker will result in voiding this warranty.  
Complete details of the warranty are available on request.  This warranty applies to units purchased 
and operated in North America.

Содержание Balston 72-730

Страница 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Страница 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Страница 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Страница 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Страница 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Страница 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Страница 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Страница 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Отзывы: