Parker Balston 72-730 Скачать руководство пользователя страница 6

Bulletin TI-72730J 

Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

6

1-800-343-4048

Description P/N

Sensor seal set 

22172

Galvanic cell (sensor) 

72695

Fuse

 13221

Cordset 

13612

 

Balston 72-730 and 72-O2730NA Oxygen Analyzer

CSA Certification Standard

 

CSA 22.2 No. 61010-1-12

IEC 61010 Installation Category

 

Category II

IEC 61010 Pollution Category

 

Degree 2

Accuracy

 

±1% Full scale calibrated span, after 30 min. stabilization

Span concentration

 

.1 to 99.9% oxygen

Response time

 

12 seconds

Display 

LED Type

Digital display range

 

00.0 to 99.9% oxygen

Min./Max. sample flow rate

 

25 scc/min. to 850 scc/min.

Min./Max. inlet pressure

 

2 psig/140 psig (0.1 barg/9.7 barg)

Min./Max. operating temperatures

 

59°F / 95°F (15°C / 35°C)

Max. Relative Humidity

 

80% to 31°C; 50% at 40°C

Altitude

 

2000 m

Sensor type

 

Galvanic cell

Sensor life

 

1 year

Required calibration frequency

 

2 weeks

Alarm outputs 

DPDT relay contacts 3 amp, 250 VAC Rating, resistive

Power requirement

120 VAC, 230 VAC, 50/60 Hz, .10 A (120 VAC), .05 A (230 VAC)

Fuse Type

 

Type T, Slow Blo Fuse, 250 V, 0.10 amp, 5 x 20 mm

Physical dimensions

 

11"h x 5"w x 4"d (28cm x 13cm x 11cm) 

Shipping weight

 

6.5 lbs. (3 kg)

*

Note: 

Main supply line voltage must be within 10% of nominal rated voltage for the generator.

System Specifications

2

3

Fuse

Replacement

to test integrity of the connection.  Repeat this procedure from step 2 if the wires release 
easily.  Replace sensor cover.

This equipment has fuses in both neutral and phase lines.  Use care when servicing.

Occasionally, one or both of the fuses in the analyzer may burn out.  The fuses are located 
in the power receptacle on the left hand side of the analyzer.  

Before servicing the fuses, 

disconnect the power cord from the local power supply and from the power entry mod-
ule.  

Both fuses should be checked each time fuse replacement is warranted.  To access the 

fuses, use a small screwdriver to remove the holder located in the power receptacle of the 
generator (see Figure 5).  Replace either one or both fuses as necessary and re-assemble.  

For continued protection against risk of fire, replace only with fuse of specified rating.

Replacement Parts

Содержание Balston 72-730

Страница 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Страница 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Страница 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Страница 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Страница 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Страница 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Страница 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Страница 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Отзывы: