Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

3

1-800-343-4048

The galvanic cell is not installed into the analyzer prior to shipment.

  It is shipped in the bag 

which contains the documentation for the product.  The tools required to install the oxygen sensor 
are a small flat head or Phillips screwdriver and wire strippers.  Install the sensor as follows:

Remove sensor cover on left side of unit.

Strip sensor connecting wires to 1/4" to 3/8" (6mm to 9mm) using wire strippers.  Install the 
sensor into the sensor holder on the left side panel of the analyzer (see Figure 3) by pushing 
firmly into place.

Insert small screwdriver into the hole adjacent to the wire connection point (see Figure 3) and 
press to open connector.

Slide the stripped wire end into the appropriate (red or black) connection port until it “bottoms 
out”.  The red wire goes to the “+” connector, and the black wire goes to the “-” connector.

Remove the screwdriver to clamp the wire into the connection port.  Pull the wire gently to test 
integrity of the connection.  Repeat this procedure from step 2 if the wires release easily.

6

 

Replace sensor cover before powering up unit.

The 72-730 and 72-O2730NA Oxygen Analyzers are preset at the factory for operation at       120 
VAC.  (

Note:

 Main supply line voltage must be within 10% of the nominal rated voltage for the 

generator.)

  

The voltage setting for the analyzer is shown through a small window on the power 

entry module on the right side of the analyzer (see Figure 5).  

Check the voltage selector set-

ting prior to energizing the analyzer.

  The selector setting should match the voltage of the local 

power supply.  (Note:  The "NA" version of the oxygen analyzer will only operate at 120 VAC.)  Plug 
the provided power cord into the power entry receptacle of the analyzer, and plug the opposite 
end into a wall outlet with earth ground protection. The North American cordset is rated for 125V, 
10 amperes, 18 AWG, SJT, with ferrite bead. Do not replace it with an inadequately rated cordset. 

The analyzer is powered on by plugging into electrical supply.  There is no power switch on 
this product.

If the voltage selector displays an input power voltage different from the local power supply, it may 
be changed using only a small screwdriver (check voltage rating of analyzer on product label on 
the bottom panel to ensure compatibility with local electrical supply).  First, use the screwdriver 
to release the cover of the power entry module on the right side of the analyzer (see Figure 5).  
Next, rotate the voltage selector until the desired input voltage is displayed in the window.  Finally, 
replace the power entry module cover.

Galvanic Cell

Voltage Selector

The 72-730 and 72-O2730NA are calibrated prior to shipment; however, Parker strongly recom-
mends re-calibrating the units prior to initial start-up.  After the initial start-up, the unit should be 
calibrated on a bi-weekly basis until a suitable schedule is determined, based upon the level of 
accuracy required by the application.

There are two methods of calibrating the 72-730 and 72-O2730NA Oxygen Analyzer:  the two 
point method and the single point method.  In the two point method,  the first point in the 
calibration range is set to zero using a zero gas (zero percent oxygen), and the second point in 
the range is set to a known percentage of oxygen using a span gas (known quantity of oxygen, 
per gas supplier) or compressed air (20.9% oxygen).  In the single point method, only one point 
in the calibration range is set, using either span gas or compressed air.  Maximum accuracy in 
oxygen concentration monitoring will be achieved if the oxygen concentration in the span gas is 
within the range of the expected oxygen concentration in the process stream.  See Figure 1 for 
calibration controls and Figure 4 for recommended calibration gas valving.

Calibration

Operation

Figure 5 - Right Side View

Содержание Balston 72-730

Страница 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Страница 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Страница 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Страница 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Страница 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Страница 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Страница 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Страница 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Отзывы: