Parker Balston 72-730 Скачать руководство пользователя страница 4

Bulletin TI-72730J 

Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

4

1-800-343-4048

4

1

CAUTION:  The Balston 72-730 and 72-O2730NA

 

Oxygen Analyzers will not provide accu-

rate readings unless calibrated on a regular basis.

  The procedure for the two point calibration 

method is as follows: 
 

1

    Isolate the 72-730 and 72-O2730NA by closing the valve between the process stream and 

the analyzer (see Figure 4).  Attach zero gas to port inlet. 

2

    Open the valve in the line from the tank of zero gas to the analyzer (gas pressure should be 

approximately equal to process pressure, if possible).

3

   Allow the zero gas to flow through the analyzer until the reading on the oxygen concentration 

display stabilizes.

4

    Adjust the zero potentiometer until the oxygen concentration display reads zero.

5

    Disconnect the zero gas from the calibration gas line, and connect a tank of span gas (pro-

cess pressure) or a source of clean compressed air.

6

    Allow the reading to stabilize, and adjust the span potentiometer until the reading on the oxy-

gen concentration display reads the known percent (span gas) or 20.9% (compressed air).

7

    Close the valve from the calibration gas line and re-open the valve between the process 

stream and the analyzer.

The procedure for the 

single point calibration

 

method

 is as follows:

1

    Isolate the 72-730 and 72-O2730NA by closing the valve between the process stream and 

the analyzer. 

2

    Connect a tank of span gas (process pressure) or a source of clean compressed air to the 

calibration gas line.

3

    Open the valve in the calibration gas line to introduce the span gas or compressed air to the 

analyzer.

4

    Allow the gas or air to flow through the analyzer until the reading on the oxygen concentra-

tion display stabilizes.

5

    Adjust the span potentiometer until the reading on the oxygen concentration display reads 

the known percent (span gas) or 20.9% (compressed air).

6

    Close the valve from the calibration gas line and re-open the valve between the process 

stream and the analyzer.

(

Note:

  Parker recommends the use of a span gas with an oxygen content between 1% and 

10%.)

The high and low limits of the integral alarm may be set anywhere between 0.5% and 25% 
oxygen, depending on the process limitations.  To set the high alarm set point, press the alarm 
set switch upward and simultaneously adjust the high potentiometer until the meter shows the 
desired high alarm set point.  To set the low alarm set point, press the alarm set switch downward 
and simultaneously adjust the low potentiometer until the meter shows the desired low alarm set 
point.

The Balston 72-730 and 72-O2730NA Oxygen Analyzers also include high and low alarm relay 
contacts located on the right side panel (see Figure 5).  The oxygen analyzer, through the use 
of the alarm relay contacts, may be used to control the process stream or remotely activate an 
alarm.  For example, a high or low oxygen concentration could signal a remote alarm, open a 
backup supply of the process stream, or close the process down for protection of downstream 
equipment or processes.  The alarm relay contacts should be wired (by a qualified electrician).  
Both the high and low oxygen alarm conditions are provided with three relay outputs: a common 
(C), a normally open (NO), and normally closed (NC).

The relay contacts are rated for 250 VAC, 5 amps resistive or 1 amp inductive load or          
24 VDC 5 amps resistive or 1 amp inductive load.  Do not exceed these values in order to 
maintain the instrument safety certification.

To eliminate the possibility of electrical shock, disconnect the power cord before wiring the 
alarm relay contacts to outside circuitry.

Alarm set points

Alarm relay contacts

Содержание Balston 72-730

Страница 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Страница 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Страница 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Страница 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Страница 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Страница 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Страница 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Страница 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Отзывы: