Cleaning the compressor housing
NOTE
Danger of device fault due to liquid penetration
If liquids get into the interior of the compressor, this may
cause a fault in the device.
• Never immerse the compressor in water.
• Never clean the compressor in running water.
• Never spray any liquids onto the compressor or the power
cord.
• If liquid does get into the compressor, it must not be used
under any circumstances. Before starting the compressor
again, contact the manufacturer or dealer.
Clean the compressor housing as follows:
• Wipe the outer surface of the
housing with a clean, damp
cloth.
4.2 Multiple patient use
Cleaning and disinfection cycles
Nebuliser (with connec-
tion tubing) and mask
Replace before each patient change
Compressor housing
Disinfect before each patient change
Air filter
Replace after 200 operating hours
(approx. 1 year)
– 31 –
Содержание PARI BOY Classic
Страница 2: ......
Страница 50: ...50...
Страница 54: ......
Страница 95: ...Air pressure limit Legal manufacturer 43...
Страница 96: ...44...
Страница 102: ...50...
Страница 106: ......
Страница 148: ...44...
Страница 154: ...50...
Страница 158: ......
Страница 210: ......
Страница 262: ......
Страница 310: ...50...