Use of the compressor in professional healthcare facilities is
limited to the inpatient wards and the intensive care unit. Use
of the compressor in areas with elevated magnetic or electrical
radiation (e.g., close to an MRI scanner) is not permitted.
During transportation and storage
Minimum ambient temperature (without
monitoring of relative humidity)
-25 °C
Maximum ambient temperature (with
relative humidity of up to 93%, non-
condensing)
+70 °C
Humidity
max. 93%
Atmospheric pressure
500 hPa – 1,060 hPa
6.2 Nebuliser
General nebuliser data
Size
7
10 cm × 10 cm × 4 cm
Weight
7
31 g to 33 g
Operating gases
Air
Minimum compressor flow
3.0 l/min.
Minimum operating pressure
0.5 bar / 50 kPa
Maximum compressor flow
6.0 l/min.
Maximum operating pressure
2.0 bar / 200 kPa
Minimum fill volume
2 ml
Maximum fill volume
8 ml
7) Without mouthpiece and mask; unfilled.
– 38 –