4
REPROCESSING IN A HOME
ENVIRONMENT
The product components must be cleaned thoroughly immedi-
ately after each use, and disinfected at least once a week.
The connection tubing cannot be cleaned or disinfected.
Dry the connection tubing after each use [see: Care of the con-
nection tubing, page 27].
The maximum operating life of the connection tubing is 1 year.
4.1 Reprocessing cycles
Nebuliser
– Clean immediately after every use
– Disinfect once per week
4.2 Processing limits
Nebuliser and accessories,
disinfection
300 reprocessing cycles,
max. 1 year
4.3 Preparation
• Remove the tube from the nebuliser.
• Pull the mouthpiece off the nebuliser.
• Remove all residual fluid from the nebuliser.
• Dismantle the nebuliser into its individual parts.
• Carefully pull the blue exhala-
tion valve out of the slot in the
mouthpiece. The exhalation
valve must still be attached to
the mouthpiece.
– 22 –